Re: [AMMRL] AVANCE I HPPR communication error

From: DrSpin via groups.io <rajanp=uw.edu_at_groups.io>
Date: Fri, 21 Mar 2025 15:23:06 -0700

Hi Beth

   Before some of the wizards weigh in with unique solutions, I will just
do a sanity check with the following :

   - Did you run the 'hppr' shell level tool and then do a '2.
   Check/Download HPPR' ? In our case, any kind of voltage fluctuation used
   to cause the downloaded firmware to be garbled though the lights will be
   lit green, where they are expected to be.
   - The communication error will certainly be one of the symptoms of the
   firmware being corrupt.
   - I will mention this about RX22 flashing red. When you do a 'gs' or
   'wobb' command, the RX22 will flash red. Way back when I took the AVance
   maintenance course, they mentioned about this gotcha and the unfortunate
   choice of repurposing this LED for 'wobb' operation !
   - Does your 'wobb' look OK ? If you don't get an RG value of the order
   of 50 to 100, then, check if you get the same problem at two different
   frequencies i.e. 1H and some X nucleus such as 13C. If you see noisy
   wobb output without the tell tale 'V' curve, then it is not one of your
   slices but the detection part and cover module. A HPPR download should
   help.
   - If you 'wobb' is passing, then most of the HPPR is vindicated. Also,
   if lock works as expected, then that is further certification that HPPR is
   firing on (mostly) all cylinders ! (I am saying mostly, because, I have
   been living with two Avance I units with 2H slices that are not upto par
   but they still work fine for locking purposes; that's a long story)
   - I think one can move to the console after that. If you suspect
   RX22, then you can start with the RG you end up by running 'rga' for a
   water sample. Should be pretty low, somewhere around 10.

 I will give the floor to smarter folks to weigh in !

And, best wishes to get this sorted out.

Best Regards
Rajan


--

Rajan K Paranji, Ph.D.
NMR Services Manager
Department of Chemistry
Bagley Hall 65 | Box 351700 Seattle, WA 98195-1700
206-685-2581 | rajanp_at_uw.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2163): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2163__;!!PvDODwlR4mBZyAb0!TKfmHX9oWuaodag5j4ffWc8AtFDY1_=
H1qt7MPOfrriFLnbvY6NXHjVeIC_aoXQRsy-KRC5OOJFLND9ntgGVoIj4G$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111836769=
/7559972__;!!PvDODwlR4mBZyAb0!TKfmHX9oWuaodag5j4ffWc8AtFDY1_H1qt7MPOfrriFLn=
bvY6NXHjVeIC_aoXQRsy-KRC5OOJFLND9ntgKlrRgTK$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Mar 21 2025 - 18:09:03 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST