It’s a long shot, but we have managed to keep our flaky two-channel VFS
alive for quite a few years by boosting the level of the output of the comb
generator (we used a TI TRF37D73 broadband amplifier module module).

It’s frustrating to know that schematics must be out there somewhere!

Best wishes

Gareth


> From: main_at_ammrl.groups.io on behalf of Lin, Penghui via groups.io
> Date: Tuesday, 4 March 2025 at 14:43
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] #agilent #Hardware #VarianAgilent

Thank you all for the replies,

I forgot to mention that this is a 4-channel VFS module and both channel 3 \
and 4 are also bad. Only channel 2 still works for now. So it seems the other
three channels share some kind of control or gating system?

I probably shouldn't use the term "easy fix". What I meant is if there is any
way we can diagnose the component inside the module to find out which one(s)
goes wrong and possibly replace it in-house.

Since John mentioned that there is no schematics, guess it is almost impossible.....

Penghui



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2105): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2105__;!!PvDODwlR4mBZyAb0!Xv-jBOkj3ri0LTVPoA_2sXOn2GJQm0=
ckdrnf28ZGuH1_kHWmkGgKXd_6JsJBDccu3NNknNlivF1zsRN-oGb9qGLn1l8h$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111440908=
/7559972__;!!PvDODwlR4mBZyAb0!Xv-jBOkj3ri0LTVPoA_2sXOn2GJQm0ckdrnf28ZGuH1_k=
HWmkGgKXd_6JsJBDccu3NNknNlivF1zsRN-oGb9qIMzBXHR$
Mute #varianagilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/m=
ain/mutehashtag/varianagilent__;!!PvDODwlR4mBZyAb0!Xv-jBOkj3ri0LTVPoA_2sXOn=
2GJQm0ckdrnf28ZGuH1_kHWmkGgKXd_6JsJBDccu3NNknNlivF1zsRN-oGb9qFy-LvVV$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!Xv-jBOkj3ri0LTVPoA_2sXOn2GJQm0ckdrnf=
28ZGuH1_kHWmkGgKXd_6JsJBDccu3NNknNlivF1zsRN-oGb9qNwU5QRI$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!Xv-jBOkj3ri0LTVPoA_2sXOn2GJQm0ckdr=
nf28ZGuH1_kHWmkGgKXd_6JsJBDccu3NNknNlivF1zsRN-oGb9qGa57D3Y$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Mar 04 2025 - 07:13:04 MST

This archive was generated by hypermail 2.4.0 : Tue Apr 01 2025 - 14:21:16 MST