Re: [AMMRL] MR400 LO switching

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 27 Mar 2024 23:41:03 +0100

Hi Scott,

are both 400MRs using a synthesizer board (to the right of the leftmost RefGen
board) or are they using a 19" VFS synthesizer module?

The synthesizer boards are known to be a source of problems. May be worthwhile
testing the other channel by swapping the RJ45 as well as BNC plugs between Ch1/2.

Did you check all voltages on the Molex plugs in the bottom right side of the
console? Any browning on any of the power supply pins?

Best regards
Bert

> Am 27.03.2024 um 18:40 schrieb Smith, Scott K <scott.smith_at_austin.utexas.edu>:
>
> I am sure nearly everyone has had to track down a missing signal.
>
> In this case the root cause in somewhere in the MR400’s LO switching
> device. The MR has a single receiver box for both channels. The LO used in
> detection is pre-selected by an external module (Mini Circuits RF switch ZASWA-2-50DRA+).
> In short, this switch is no longer passing through high band LO to the receiver
> making it appear that there is no signal. It looks like the LO for low band
> channel is passively allowed through, so one gets the curious result that a carbon
> spectrum with full proton decoupling can be acquired while nothing not even tuning
> works on the high band channel. All the transmitter pathways checkout as do the
> LO’s themselves. Proton work if I manually swap the LO input cable directly
> to the receiver.
>
> Are these mini circuit devices known to just quit working? Or might there be a
> problem with the power-supply and TTL control lines feeding the switch?
>
> I now have two MR’s with this error.
>
> I would appreciate any suggestions or help I can get at this point.
>
> Cheers,
> Scott
>
>
>
> Scott K. Smith, Director of NMR Spectroscopy
> The University of Texas at Austin | Chemistry | 832-294-1928 | utexas.edu
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1098): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1098__;!!PvDODwlR4mBZyAb0!VUMHEDpsRbn-gMBN4GHpSRjpQqUUW3=
sYcLNd4ExFhUioYXqHTAfJPiJ_NdTOLWHaOqSSH7gyZfuY_yfuSKBLAA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105182325=
/7559972__;!!PvDODwlR4mBZyAb0!VUMHEDpsRbn-gMBN4GHpSRjpQqUUW3sYcLNd4ExFhUioY=
XqHTAfJPiJ_NdTOLWHaOqSSH7gyZfuY_yf6Tmystw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Mar 27 2024 - 16:27:27 MST

This archive was generated by hypermail 2.4.0 : Fri Apr 12 2024 - 18:03:18 MST