Re: [AMMRL] Parts for Bruker Avance console #Bruker #Hardware

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 27 Mar 2024 23:16:03 +0100

Hi Anne,

the green box should normally be relatively easy to repair - it's a (shockingly
simply built out of only one single diode and two electrolytic capacitors…)
rectifier, typically the capacitors inside overheat and blow up. Replace the caps,
maybe re-solder all connections (I had one which essentially unsoldered itself)
and it should normally work again.

Best regards
Bert

> Am 26.03.2024 um 20:24 schrieb Anne Carroll <carrol50_at_miamioh.edu>:
>
> Hello,
> I am hopeful that someone here has an old Avance console they are keeping
> for parts and are willing to scrounge a few elements for me from the safety
> box at the back. I have tracked the issue with our 500 MHz SS NMR to the
> narrow switch (EH-04-11) attached beside the larger relay (ESB-63-40). I am
> also missing the green Temporizer that should go next to the transformer (TM15/24).
> Please let me know if you have these components to spare, EH-04-11 and Temporizer.
>
> Thank you,
> Anne Carroll
> --
> Anne M. Carroll, Ph.D. (she/her)
> NMR/MS Instrumentation Specialist
> Department of Chemistry and Biochemistry
> Miami University
> Oxford, Ohio USA
> 513-529-3163 Office
> 678-637-6739 Cell
> carrol50_at_miamioh.edu <mailto:carrol50_at_miamioh.edu>
>
> <Safety Box 3-14-24-edited.jpg>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it
MestreLab NMR Software (http://www.mestrelab.com
Barthel HF Amplifiers (http://www.barthel-hf.de


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1097): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1097__;!!PvDODwlR4mBZyAb0!Q_1DK49c7daY6p2JNYjcIXvlvxcJgl=
1U6-S_v9YZm1zsloMTxbD7UNPit5dBpct3qCbESuisscND9meUtFX1iA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105175496=
/7559972__;!!PvDODwlR4mBZyAb0!Q_1DK49c7daY6p2JNYjcIXvlvxcJgl1U6-S_v9YZm1zsl=
oMTxbD7UNPit5dBpct3qCbESuisscND9mf8dLKTfQ$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!Q_1DK49c7daY6p2JNYjcIXvlvxcJgl1U6-S_v9=
YZm1zsloMTxbD7UNPit5dBpct3qCbESuisscND9mfcfg-D7A$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!Q_1DK49c7daY6p2JNYjcIXvlvxcJgl1U6-=
S_v9YZm1zsloMTxbD7UNPit5dBpct3qCbESuisscND9mfQ3J7PNg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Mar 27 2024 - 16:27:13 MST

This archive was generated by hypermail 2.4.0 : Fri Apr 12 2024 - 18:03:18 MST