Re: [AMMRL] problems with pulsecal --> One way valves for helium exhaust

From: Jeff Walton via groups.io <jhwalton=ucdavis.edu_at_groups.io>
Date: Tue, 5 Mar 2024 20:06:11 +0000

This was our solution at UC Davis……

Jeff

On Feb 22, 2024, at 9:17 AM, Jones, David (Pharm) via groups.io wrote:

On our older Oxford magnets we have installed these higher flow valves from
generant. They have cracking pressure of 10 mBar (0.15 psi). They just need
a KF25 to NPT adapter that we get at Loco Science. They are in line and
upstream of our diverter for switching between routine operation and fills. o
They are brass and significantly heavier than the valves from Bruker.


https://www.generant.com/product/check-valve-cv/

Check Valve (CV) – Manufacturer of Relief Valves, Check Valves, Needle Valves,
Pressure Regulators, LTAP and Cryogenic Products

David Jones

Associate Professor
Dept. Of Pharmacology
University of Colorado
School of Medicine
Tel (303)-724-3600

On Feb 22, 2024, at 8:59 AM, Mike Lumsden via groups.io wrote:


Hello - regarding the one-way valve discussion, we also have a mixture here
at Dalhousie (all 4K magnets). Older Bruker magnets have the brass lower-flow
valves which we remove prior to a fill. These are from Circle Valve which,
in our case, look to have a FNPT-->KF flange on one end and a barb fitting on
the other. The cracking pressure is listed at 10 mbar A recent Ascend magnet
was fitted with Z52395 which we leave on during fills. I read somewhere the
cracking pressure for this valve is 15 mbar. To make matters more interesting,
v1 of the Bruker document "Helium Recovery for NMR Systems" listed the exhaust
valve for 4K magnets as part number Z58140 (the "installation only" valve) but
in v2 this was updated to Z52352.

In terms of capturing helium, I too have been wondering if the older magnets
with the Circle Valves should be updated to the higher flow style OR if I should
simply remove them from the back of the magnets and tee them into the low flow
pathway. To replace them, its not clear to me what the proper part number is as
there are three that I know of floating around (Z52395, Z58140, or Z52352).

Some clarity on these different valves would be most helpful!

Thanks,
Mike


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1052): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1052__;!!PvDODwlR4mBZyAb0!VuzUbN63DRJtddYaBMu1g1hHMa1_xa=
l_iCGUipqBVyuPJciobnf9WGCJefhTDR_YrzDhSYQgcuAeIBnFFyOo2mtUB2Ah$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104493541=
/7559972__;!!PvDODwlR4mBZyAb0!VuzUbN63DRJtddYaBMu1g1hHMa1_xal_iCGUipqBVyuPJ=
ciobnf9WGCJefhTDR_YrzDhSYQgcuAeIBnFFyOo2hUzggjY$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Mar 05 2024 - 12:06:18 MST

This archive was generated by hypermail 2.4.0 : Fri Apr 12 2024 - 18:03:18 MST