Re: [AMMRL] TopSpin installation & CodeMeter issues

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Fri, 12 Jun 2026 15:34:51 +0000 (UTC)

Hi Geoff,

This has been submitted as a ticket internally but I don't have any idea if
there's a timeline for it to be fixed.  It's been going on for a couple o
f months now but there's not a lot of information that I can see in the
bug report.  

For what it's worth, it appears to be due to the codemeter website being an
http site instead of https.  It tries to run something on the computer
and if it were an https site, chrome or Edge could give you the option to
let it run or not.  Since it's not an https site, it just doesn't run
and you never get the option to allow it.  At least, that's what I
gathered from the bug report.  There may also be a ticket open with the
codemeter support people but I don't have access to that information. 
You can still use chrome or Edge to install the license manually--there are
a few more steps but it's not super painful.  As long as you make sure
to upload the receipt file at the end to avoid future problems.  You
can even do that from the spectrometer computer if it's got internet access
and you don't want to install firefox for some reason.  

Cheers,Mike

On Friday, June 12, 2026 at 08:11:25 AM MDT, Akien, Geoffrey via groupsio wrote:

Hi all,

I didn’t see this problem when I was setting up deployment of 4.5.0
on university W11 machines back in October 2025, but now I’m seeing
the same issues with Chrome and Edge, and Firefox fixing everything. This
seems to be a problem with the current version of WebDepot v23.03.554.500.ws4.
As far as a I can tell, the specific version of TopSpin is irrelevant (tested
both 4.5.0 and 5.0.0) – it’s whatever’s on Wibu’s
servers that seem to be the issue.

Mike do you know if this has been submitted as a proper ticket internally?
Do we have a timeline for this getting fixed? I don’t see how not
supporting 75% of the installed browser market is winning strategy…

Thanks

Geoff


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3515): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3515__;!!PvDODwlR4mBZyAb0!R_odCPv7JqJEoZZ1VJUU6xw6ZoPD9o=
eJ_-FakaEIM3Oe-hn-ij9T95UOdyqcNuStmhCuOaTPZqQHvVGct9ZuAhbl6XM5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/119550449=
/7559972__;!!PvDODwlR4mBZyAb0!R_odCPv7JqJEoZZ1VJUU6xw6ZoPD9oeJ_-FakaEIM3Oe-=
hn-ij9T95UOdyqcNuStmhCuOaTPZqQHvVGct9ZuAk5UFO7b$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 12 2026 - 08:44:23 MST

This archive was generated by hypermail 2.4.0 : Tue Jun 30 2026 - 16:28:21 MST