Hi,

We have a similar experience with an old Inova 400 and Oxford magnet. 
In our case, the helium level is read properly in slow mode, but when
switched to fast mode it immediately jumps to 100% and stays there until
the next day even after switching back to normal mode.  So it is useless
during a helium refill.

Regards,
Eugenio

Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924

On 6/27/25 8:44 AM, foglein via groups.io wrote:

> I've been lurking on this group for many years, but I believe this is
> my first time asking a question.
> We have an Oxford AS400 magnet as part of a Varian Mercuryplus system.
> Starting with the last helium refill in April 2025, the helium level
> probe no longer reads an accurate liquid helium level. When switched
> on for a nitrogen fill, the helium level initially reads ~91-94%, and
> switching to the 'fast' rate doesn't register a lower reading, which
> would be more accurate to the actual level inside the magnet. Using
> historical data, I am performing a helium refill today, so there is no
> fear of quenching. Reading through the manual, I could not find a
> procedure for rectifying this issue. I would appreciate anyone's
> recommendations for resolving this issue, especially if you have
> experienced this issue yourself, and how you resolved it.
> Thanks in advance,
> Jon
>
>
> Mr. Jonathan Foglein
> Instrument Coordinator/
> Chemical Hygiene Officer
> Rowan University
> Department of Chemistry & Biochemistry
> Science Hall, Room 345
> 201 Mullica Hill Road
> Glassboro, NJ 08028
> Phone: 856-256-5456
> Email:_foglein_at_rowan.edu <mailto:foglein_at_rowan.edu>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2463): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2463__;!!PvDODwlR4mBZyAb0!TiYGf7THpX0ieqm687g2ic_0GSry3q=
xaHnuP5NFyZ4xHjWt26KhOdwWtpiY-r-wd7VhL3UAqxyq1LkHDo3IwkYRxMdc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113860576=
/7559972__;!!PvDODwlR4mBZyAb0!TiYGf7THpX0ieqm687g2ic_0GSry3qxaHnuP5NFyZ4xHj=
Wt26KhOdwWtpiY-r-wd7VhL3UAqxyq1LkHDo3Iw1R_rMZM$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!TiYGf7THpX0ieqm687g2ic_0GSry3qxaHn=
uP5NFyZ4xHjWt26KhOdwWtpiY-r-wd7VhL3UAqxyq1LkHDo3IwoIWih-c$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!TiYGf7THpX0ieqm687g2=
ic_0GSry3qxaHnuP5NFyZ4xHjWt26KhOdwWtpiY-r-wd7VhL3UAqxyq1LkHDo3IwMLWc9Kk$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 27 2025 - 07:39:17 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST