Re: [AMMRL] Bruker Avance BSMS scan trigger

From: Shore, Jay via groups.io <jay.shore=sdstate.edu_at_groups.io>
Date: Thu, 26 Jun 2025 02:11:19 +0000

Thank you very much.

Am I correct in assuming that this is a TTL line?

Thank you again.

Jay

________________________________
> From: main_at_ammrl.groups.io on behalf of Spin-Doc via groups.io
> Sent: Wednesday, June 25, 2025 8:50 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Bruker Avance BSMS scan trigger

Hi,

that's unfortunately a bit complicated cable it is a male (usually plugs are
female) SMB plug on the BSMS side going into a thick harness and connects
to connector C, bottom left on the AQS FCU board. If lucky you can extract
that pin (after double-checking) with the pin extractor tool you may have
in one of the blue toolkit boxes - at least older ones had one of them.

I have a feeling that cutting off the golden SMB plug and soldering a new one
onto the existing cable is probably the easiest way to repair this the SMB plug
I'm looking at is even screwed instead of crimped so you may be able to re-use it

Cheers

Bert

-> Sent via mobile <-

Spin-Doc
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Web: http://www.spin-doc.net
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF

Am 26.06.2025 um 02:30 schrieb Shore, Jay via groups.io:

The cable to our BSMS "scan trigger" port broke. It is on the board that has
the helium level and sample control in the BSMS unit of our Avance spectrometer.
I was wondering if anyone knew where the other end of the cable went. I was
thinking that just replacing the cable might possibly be the easiest fix.

I would appreciate any helpful suggestions.

Jay


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2457): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2457__;!!PvDODwlR4mBZyAb0!Vg8BcKUp-NksyXMG6lWuFUhTEugN2N=
1hiTiMFf_oploK96_8Hz_KOFCcNlakVe9hzg9TwYCz4E6ki_aKPsFv6ZyOPn5V$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113836192=
/7559972__;!!PvDODwlR4mBZyAb0!Vg8BcKUp-NksyXMG6lWuFUhTEugN2N1hiTiMFf_oploK9=
6_8Hz_KOFCcNlakVe9hzg9TwYCz4E6ki_aKPsFv6UbSlRiU$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jun 25 2025 - 19:30:08 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST