Re: [AMMRL] Bruker Avance BSMS scan trigger

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Thu, 26 Jun 2025 03:50:58 +0200
Hi,

that’s unfortunately a bit complicated cable… it is a male
(usually plugs are female) SMB plug on the BSMS side going into a thick harness
and connects to connector C, bottom left on the AQS FCU board. If lucky you can
extract that pin (after double-checking) with the pin extractor tool you may have in
one of the blue toolkit boxes - at least older ones had one of them. 
I have a feeling that cutting off the golden SMB plug and soldering a new one onto the
existing cable is probably the easiest way to repair this… the SMB plug I‘m looking
at is even screwed instead of crimped so you may be able to re-use it.

Cheers 
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF



Am 26.06.2025 um 02:30 schrieb Shore, Jay via groups.io:

The cable to our BSMS "scan trigger" port broke. It is on the board that has the helium level and sample control in the BSMS unit of our Avance spectrometer. I was wondering if anyone knew where the other end of the cable went. I was thinking that just replacing the cable might possibly be the easiest fix.
I would appreciate any helpful suggestions.
Jay


_._,_._,_

Groups.io Links:

You receive all messages sent to this group.

View/Reply Onl= ine (#2456) | Reply to Gro= up | Reply to Sender | Mute This Topic | New Topic
Your Subscription | Contact Group Owner |

_._,_._,_
Received on Wed Jun 25 2025 - 18:59:28 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST