From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Tue, 3 Jun 2025 17:20:53 +0000

We have been told that Multiple Solvent Suppression should be possible in
automation and were directed at the LC-NMR manual. The manual does refer to
performing these experiments in automation, but I have not been successful.
Hoping that someone here has been able to get this to work.

We would like to suppress at least 3 peaks. Procedurally, you would first
run a proton, integrate targets peaks and save a .reg file and then run the
solvent suppression. There are two variations - a MULTIPRESAT and a LC1DC
WGPPS parameter sets. Both use noesy shaped pulses. I haven't been successful
with either from within ICON or without. They rely on an au program au_lc1d.

I did get it to work in topspin using an older wet1dc program.

One issue seems to be that while I define one set of frequencies in the .reg
file, it chooses a 3rd peak I didn't select as shown in the image below.



Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2368): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2368__;!!PvDODwlR4mBZyAb0!XcAtKf-yzdk9WvFN2zAP0xaijbyLLK=
NvT2bNUIwGMnVdESR2xH2KLPuFy-d4Lvu1zFkok79KB3F2aN5bHvHttsT_$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/113451547=
/7559972__;!!PvDODwlR4mBZyAb0!XcAtKf-yzdk9WvFN2zAP0xaijbyLLKNvT2bNUIwGMnVdE=
SR2xH2KLPuFy-d4Lvu1zFkok79KB3F2aN5bHnwiTXxf$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






image.png
(image/png attachment: image.png)

Received on Tue Jun 03 2025 - 10:21:33 MST

This archive was generated by hypermail 2.4.0 : Tue Jul 01 2025 - 15:46:21 MST