Re: [AMMRL] Agilent 7600-AS error -3128 encoder battery down

From: Sheahan, Charles Andrew <Charles_Sheahan_at_hms.harvard.edu>
Date: Fri, 28 Jun 2024 13:24:48 +0000

Hey Dan,

If memory serves, Cal_PP should have been installed on the robot flash memory
when the system was set up by the Agilent engineer. If you haven't tried to
follow the instructions to access the internal page I would do so just to
check things out. In the meantime, I'll check around here and see if I
can dig up a copy of the CD.

I also don't know if you've ever had any dealings with Deane Alexander, but
he was one of the field engineers Agilent recently let go... He was a huge
help back in the day when I was having issues with my 7600AS. If you can
contact him (although without the official Agilent email that might be hard),
he might know something.

-C

Charles A. Sheahan
NMR Manager
Harvard Medical School
617-432-6025
________________________________
> From: main_at_ammrl.groups.io on behalf of Holmes, Daniel
> Sent: Friday, June 28, 2024 9:13 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Agilent 7600-AS error -3128 encoder battery down

Thanks, Charlie,

I will reach out to Jonathan and likely NMR associates as well. I found both
the user manual and the installation manual which walks through the reset
procedure. Worryingly, the installation manual says I need to run Cal_PP
that is supplied on the Guidance Controller System Software CD…


Dan

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1381): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1381__;!!PvDODwlR4mBZyAb0!X9OEW6bC3mmsrohJSrjqktzwBEKNIg=
Kgk0oot6QfsxE_O7XGHcCWYEo9YT2ZElU3bBvy_R32tMyfRUIFGtV6o8xP51lRCAEK$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106921425=
/7559972__;!!PvDODwlR4mBZyAb0!X9OEW6bC3mmsrohJSrjqktzwBEKNIgKgk0oot6QfsxE_O=
7XGHcCWYEo9YT2ZElU3bBvy_R32tMyfRUIFGtV6o8xP56-o4d6a$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 28 2024 - 06:24:54 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST