Re: [AMMRL] Flirting with the 5 gauss line #HeliumRecovery

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Fri, 21 Jun 2024 15:04:52 +0000

Hello Luke,

You say that the old 400MHz magnet is “… not decommissioned
yet.”
Implicit in that statement is that the old 400MHz magnet will be decommissioned
soon.

My suggestion: put a stop to the installation of the storage tank in question
until the old magnet is de-energized. Could your recovery system operate
with two storage tanks temporarily? Then, once the old magnet is removed,
the third storage tank added?

Too many things can go horribly wrong when you have a big stray field and large,
heavy magnetic objects, as well was people working with heavy metal tools.
And this doesn’t even consider how difficult/impossible it will be to
record any useful data using the 400MHz.

Good luck.
Bob.

> From: main_at_ammrl.groups.io On Behalf Of Fulton, Luke via groups.io
> Sent: June 21, 2024 10:45 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Flirting with the 5 gauss line #HeliumRecovery

Hi All,

I'm seeking the collective wisdom regarding a site planning pickle.

We're replacing our old 400 MHz magnet/console later this summer. The grant
also included a full helium recovery system with three medium pressure storage
tanks that we're in the midst of installing. I think it was assumed the new
magnet would be installed before the recovery system, and that didn't happen.
The old magnet hasn't been decommissioned yet and it's unshielded. The 5 gauss
line extends ~7ft from the magnet, and the plan has one of the tanks crossing
it by a full foot. Each tank is 8.5ft ht, 3ft dm, magnetically susceptible
(steel?), and ~400 lbs. We have to permanently bolt down all three next week.
I can rearrange the plan, but it would be a tragedy for the lab layout, and
I'd never hear the end of it.

Does it seem like an acceptable risk to put that much metal inside the stray
field?

My building manager thinks we could slide the magnet a foot or two out of the
way with johnson bars. I've only ever heard the direst of warnings about
trying to move a magnet, and I'm skeptical we could even pull it off if we
tried.

Kind regards,
Luke

Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
nit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu<mailto:fko24003_at_uconn.edu>
(alias): luke.fulton_at_uconn.edu<mailto:luke.fulton_at_uconn.edu>
mobile: (603) 953-5275
Office: (860) 486-4069



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1363): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1363__;!!PvDODwlR4mBZyAb0!SX_PC1mIDW8GUcpPw7a6u-k57N1kQh=
ZdiLizNHQaak8xURKBAKAdeVI7GOWVegi4fc7r521-HSILAUNx$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106800203=
/7559972__;!!PvDODwlR4mBZyAb0!SX_PC1mIDW8GUcpPw7a6u-k57N1kQhZdiLizNHQaak8xU=
RKBAKAdeVI7GOWVegi4fc7r521-HUq3KUQq$
Mute #heliumrecovery:https://urldefense.com/v3/__https://ammrl.groups.io/g/=
main/mutehashtag/heliumrecovery__;!!PvDODwlR4mBZyAb0!SX_PC1mIDW8GUcpPw7a6u-=
k57N1kQhZdiLizNHQaak8xURKBAKAdeVI7GOWVegi4fc7r521-HV9njTQA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jun 21 2024 - 08:04:57 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST