Dear Karl,

I experienced similar problems regularly during summer at my old lab. My
experience is that at around 28°C / 83 F the first boards gave up and
produced errors, though they did not suffer permanent damage and came back
to life once the temperatures were lower. Regardless, I would encourage
you to shut down the spectrometers before this point is reached.

Bests,

Christian


Dr. Christian Adam
Analytics and Material Physics
Polymer, Quantitative and Heteronuclear NMR

Phone: +49 621 60-58670, Mobile: +49 172 6352576, Fax: +49 621 66-58670,
Email: christian.k.adam_at_basf.com
Postal Address: BASF SE, RGA/AS - B009, Carl-Bosch-Strasse 38, 67056
Ludwigshafen am Rhein, Deutschland
BASF SE, Registered Office: 67056 Ludwigshafen, Germany
Registration Court: Amtsgericht Ludwigshafen, Registration No.: HRB 6000
Chairman of the Supervisory Board: Kurt Bock
Board of Executive Directors:
Markus Kamieth, Chairman;
Dirk Elvermann, Michael Heinz, Anup Kothari, Stephan Kothrade, Katja Scharpwinkel

Information on data protection can be found here:
https://www.basf.com/global/en/legal/data-protection-at-basf.html


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1358): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1358__;!!PvDODwlR4mBZyAb0!Q3YmaoIzHfHV3u3XTy0ac5Qu9gYASP=
SYzxaV5r3uG8yQPvsm7AQKVBywaWOJ6xg8ylbwnh0zMF25VEys1vKPokvJSEdD$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106759058=
/7559972__;!!PvDODwlR4mBZyAb0!Q3YmaoIzHfHV3u3XTy0ac5Qu9gYASPSYzxaV5r3uG8yQP=
vsm7AQKVBywaWOJ6xg8ylbwnh0zMF25VEys1vKPonNPsV7m$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jun 20 2024 - 11:09:40 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST