Re: [AMMRL] Maximum room temperature for running NMR

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Wed, 19 Jun 2024 14:36:11 +0200

Hi,

usually the specificatin is around max. 26 deg C but I have some customers
in Europe which occasionally struggle to keep the room temperature under 30…
But above 30 I would rather switch it off.

At least I would recommend to check ALL fans in the consoles - the Neo isn't
that old so should be OK, but the Inova is way older and may have some dead
or struggling fans. The most critical part is the AMT RF amplifier - please
keep an eye on that one specifically. Make sure it isn't blocked by dust.
Open or remove front doors to increase airflow a bit. Maybe you can place
a portable AC in front of each console…?

Best of luck, stay healthy!
Bert

> Am 19.06.2024 um 04:58 schrieb kbishop_at_maine.edu:
>
> Dear AMMRL Colleagues,
>
> In central Maine as in other parts of the Northeastern US we are experiencing
> a heat wave. The central air conditioning for our building is down.
> We have an INOVA 400 and Bruker Avance Neo 500 in our NMR lab. Four portable
> AC units have been moved into the NMR lab to try to keep the lab cool.
> But they are struggling to keep the room cool. I need to know at what temperature
> I should shut down the NMR spectrometers.
>
> Thank you.
>
> Karl
>
> Karl D. Bishop, Ph.D.
> Research Associate & Instrumentation Specialist
> Department of Chemistry
> 146 Aubert Hall
> University of Maine
> Orono, ME 04469
> 207-581-1195
> kbishop_at_maine.edu <mailto:kbishop_at_maine.edu>
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1357): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1357__;!!PvDODwlR4mBZyAb0!Ta4xF-0icXgRBKEy3ip_f4RUblRWSn=
xZUZzGRD8iWj44VxFTEJ7Pg6xLAro0mDJZ0gLGr78TNBqxx0jG_bR3Fg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/106759058=
/7559972__;!!PvDODwlR4mBZyAb0!Ta4xF-0icXgRBKEy3ip_f4RUblRWSnxZUZzGRD8iWj44V=
xFTEJ7Pg6xLAro0mDJZ0gLGr78TNBqxx0iQeqpPoQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Thu Jun 20 2024 - 11:09:16 MST

This archive was generated by hypermail 2.4.0 : Mon Jul 01 2024 - 12:45:11 MST