Hi all,
I'll second what Bob said--HVAC people are a great resource in this case.
This doesn't sound like they typical failure we see from BCUs--frequently they
either stop cooling or just fail to power up altogether. My first guess would
probably be a failing capacitor and this is definitely something your HVAC guys
should be able to help with. Because it's tripping the breaker and not blowing
the fuses in the unit that does seem to indicate a capacitor issue to me.
What's the rating of the breaker that's being tripped?
Cheers,Mike
On Friday, July 17, 2026 at 10:00:01 AM MDT, Bob Berno via groups.io wrote:
Hello Luke,
I have had many issues with our BCU-2 chillers over the years. In fact,
very recently I was able to scavenge enough spare parts from two failed units
to get one of them operational again.
I suggest that you talk to the people in charge of maintaining your building
HVAC systems. They have been very helpful in diagnosing the issues with
our faulty BCU (1 and 2) chillers over the years. And they can source
replacement switches and relays, which almost always are the root cause of
the failure.
Good luck,
Bob.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3589): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3589__;!!PvDODwlR4mBZyAb0!WxxDgh95qtZaEwWk4cGThXh-42qIOU=
xvdrIQyjGInY7mwhR36fJ3lemUKV0wFLZdf-k9JcG4fE6u5WKsLwwvAHVlCqQL$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120316849=
/7559972__;!!PvDODwlR4mBZyAb0!WxxDgh95qtZaEwWk4cGThXh-42qIOUxvdrIQyjGInY7mw=
hR36fJ3lemUKV0wFLZdf-k9JcG4fE6u5WKsLwwvADaWdT6q$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Jul 17 2026 - 09:09:38 MST