I have had many issues with our BCU-2 chillers over the years. In fact,
very recently I was able to scavenge enough spare parts from two failed
units to get one of them operational again.
I suggest that you talk to the people in charge of maintaining your
building HVAC systems. They have been very helpful in diagnosing the
issues with our faulty BCU (1 and 2) chillers over the years. And they
can source replacement switches and relays, which almost always are the
root cause of the failure.
Good luck,
Bob.
On 2026-07-17 10:57 a.m., Fulton, Luke via groups.io wrote:
> Hi All,
> This week our BCU II started experiencing electrical problems. Roughly
> 5 minutes after turning on cooling the circuit breaker trips and cuts
> power. This is reproducible and happens even when given its own
> circuit without other devices. It does not matter whether operating in
> remote mode when turning on VTU or manually selecting cooling strength
> at the BCU II itself. The Status LED consistently blinks Green from
> start right up to power loss. I presume the BCU II is drawing an
> excessive load, but can't explain why. There were not any surges or
> interruptions to building power coinciding with when it first appeared
> (afaik), and this is the only device affected. In case it matters: NEO
> console, TBO probe, TS 4.4
> -I've checked the dust filter. It was cleaned recently and has a long
> ways to go before clogging enough to choke air flow.
> -I set the BCU II to flush for 24+ hours in case of ice blocks. No changes observed.
> It's less than 2 years old. Fingers crossed the problem doesn't require sending it out for repair.
> Anyone have any suggestions on what to try next?
> Kind regards,
> Luke Fulton
>
> Luke Fulton, PhD â=
> CHEM BLDG R003
> NMR Core Facility Director
> Unit 3060
> COR2E & Department of Chemistry â€=
> 55 N Eagleville Road
> University of Connecticut
> Storrs, CT 06279
>
--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.
McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3588): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3588__;!!PvDODwlR4mBZyAb0!QxfYpyIlNo_YYP69X7LFwKlapBMPdC=
KukIhW_W659lqMGeemFsZEURYeCEOiwRxQ4sLDMsbVdRk1kGV-LLEjdmFv4eql$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120316849=
/7559972__;!!PvDODwlR4mBZyAb0!QxfYpyIlNo_YYP69X7LFwKlapBMPdCKukIhW_W659lqMG=
eemFsZEURYeCEOiwRxQ4sLDMsbVdRk1kGV-LLEjdjfgb3gb$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Jul 17 2026 - 09:00:02 MST