Hi Andrei,
The MAS Display has a recording tab where you can configure recording of the
supply pressure if you like. You just have to tell it what the sampling
interval is and which parameters you want to record. The other option
would be to go into the MAS/3 service webpage where you can save log files,
and save the logs periodically. I don't remember what the form at of
those logs is, though, so I believe that the recording tab gives the most
similar solution to your previous solution. That would produce a log file
that you can monitor remotely.
Cheers,Mike
On Wednesday, July 15, 2026 at 10:45:16 AM MDT, a.gurinov via groups.io wrote:
Hi all, On the older systems equipped with MAS2, we used a simple and
dirty Bash script that connects remotely via SSH to monitor the main gas
supply pressure by extracting it from the MasDebugServer.log file in TopSpin
curdir folder.
Our newer NEO console with MAS3 is connected to a separate gas line, and we
were hoping to use the same approach. However, we have not been able to find
any pressure data that is accessible via SSH. Do you know of any source
where it is available, or could you suggest an alternative?
Thanks in advance
Best regards,Andrei NMR Spectroscopy
Bijvoet Centre for Biomolecular Research
Utrecht University
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3567): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3567__;!!PvDODwlR4mBZyAb0!TFu6spgyTIAJGnpg1YIR4LkSckGCVH=
6xMZv2zcc2zzpergOBgPX0HOVKziazlgxo37PpsXpViK8HubdQjeRXDrBZVAkH$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120285105=
/7559972__;!!PvDODwlR4mBZyAb0!TFu6spgyTIAJGnpg1YIR4LkSckGCVH6xMZv2zcc2zzper=
gOBgPX0HOVKziazlgxo37PpsXpViK8HubdQjeRXDgz7QKyk$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Wed Jul 15 2026 - 10:22:07 MST