Hi Piotr,
Both of those can be done in the User Manager page of the icon configuration.
In the upper right you'll see the experiment list. Select whichever
experiments you want to make unavailable and then click Delete in the window
that opens (this doesn't delete the parameter sets, just removes them from
the experiment list). Once you've done that, click the icon to the left
of that window that looks like a couple of sheets of paper and you can select
which users you want to make the change for and then click update. And
then click the file menu in the upper left and click save.
To edit the parameters available you can add them in the "User Specific
Parameters/commands" field. Click the edit icon to the left of that
field and it will pop up a window that will let you find and add parameters.
After that you can do the same as above to copy that to specific users and
then save in the file menu. After you've made these changes you probably
need to restart icon to apply the changes.
A lot of this stuff is covered in the ICON configuration manual. To find
this, click the help icon in topspin (either top center or top right) and
select manuals in the list that appears. The "ICON NMR Automation Interface"
manual has a lot of useful information.
Cheers,Mike
On Friday, July 10, 2026 at 09:47:22 AM MDT, Piotr Dobrowolski via groups.io wrote:
New Topspin 3.6.5 and 4.5.0 administrator.
Two IconNMR questions:
1. How can I hide experiments from the list of the available experiments
( I need to do it for all users)?
2. For a given experiment, how can I add more parameters to the "=" list?
Thank you,Piotr
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3558): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3558__;!!PvDODwlR4mBZyAb0!XE1g7KEzosuade_L1VQcqPaT4kdQ3m=
BiUvSXJRfBa5HNSdsrqfczMcxgewd7OBeRe_tglWpj9a-v0TCyQlJw46QXwZ_i$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120209024=
/7559972__;!!PvDODwlR4mBZyAb0!XE1g7KEzosuade_L1VQcqPaT4kdQ3mBiUvSXJRfBa5HNS=
dsrqfczMcxgewd7OBeRe_tglWpj9a-v0TCyQlJw4zO_Ab2V$
Mute #bruker:
https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!XE1g7KEzosuade_L1VQcqPaT4kdQ3mBiUvSXJR=
fBa5HNSdsrqfczMcxgewd7OBeRe_tglWpj9a-v0TCyQlJw4xF6rLtW$
Mute #nmr_liquids:
https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!XE1g7KEzosuade_L1VQcqPaT4kdQ=
3mBiUvSXJRfBa5HNSdsrqfczMcxgewd7OBeRe_tglWpj9a-v0TCyQlJw4xveH3xx$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Jul 10 2026 - 10:28:47 MST