Piotr,

you can edit the list of experiments and available parameters in the
User Manager section of the IconNMR Configuration.  Select one user,
click the Edit icon next to the experiment list or User Specific
Parameters list, and make the changes like removing the selected
experiment/parameter  or adding an experiment/parameter. Save it for
the selected user, then use the "Copy to other users" icon above the Edit
icon to copy the setting to other users (you can select multiple users).

If all you want to do is delete many experiments, it may be faster to
manually edit one user configuration file that is located in
<TSDIRECTORY>/conf/instr/spect/inmrusers/<username> in Topspin 3.x,
replace remote_spect for spect in Topspin 4.x.

I would not recommend adding experiments or parameters that way as there
is no check for typos. Then go back to ICONNMR and copy that user
configuration to all users.

In order for new users to inherit the same settings, I copied one user
file to a template file, that gets copied by a homemade script to the
new user file when I create a new user (I am using Linux). That way I do
not need to modify a new user from Bruker defaults after the fact.


Holger

On 7/10/2026 10:08 AM, Piotr Dobrowolski via groups.io wrote:
> New Topspin 3.6.5 and 4.5.0 administrator.
> Two IconNMR questions:
> 1. How can I hide experiments from the list of the available
> experiments ( I need to do it for all users)?
> 2. For a given experiment, how can I add more parameters to the "=" list?
> Thank you,
> Piotr
>


--
Frank H. Foersterling
Instrumentation Specialist
Chem-B87
Department of Chemistry
University of Wisconsin-Milwaukee
414-229-5896(Lab) 414-229-5035 (Teams)
holger_at_uwm.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3557): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3557__;!!PvDODwlR4mBZyAb0!X4vXBqd4l1ya181Se6ICzcrIsoqgsl=
4p0wA8XEV1M_TBtlzRClZPzCWktGm8rDz56rtCBo0tH-e3nBvwzdNn9mZ-Sw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120209024=
/7559972__;!!PvDODwlR4mBZyAb0!X4vXBqd4l1ya181Se6ICzcrIsoqgsl4p0wA8XEV1M_TBt=
lzRClZPzCWktGm8rDz56rtCBo0tH-e3nBvwzdOQ1rSuKQ$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!X4vXBqd4l1ya181Se6ICzcrIsoqgsl4p0wA8XE=
V1M_TBtlzRClZPzCWktGm8rDz56rtCBo0tH-e3nBvwzdMrfOuYiw$
Mute #nmr_liquids:https://urldefense.com/v3/__https://ammrl.groups.io/g/mai=
n/mutehashtag/nmr_liquids__;!!PvDODwlR4mBZyAb0!X4vXBqd4l1ya181Se6ICzcrIsoqg=
sl4p0wA8XEV1M_TBtlzRClZPzCWktGm8rDz56rtCBo0tH-e3nBvwzdP-01_lYg$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jul 10 2026 - 10:28:40 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST