Re: [AMMRL] J Young Tubes getting stuck in Bruker 400 with CASE autosampler

From: Ulrich via groups.io <Ulrich.Haunz=uni-konstanz.de_at_groups.io>
Date: Tue, 07 Jul 2026 10:46:02 +0200

Dear Justin,

we use also these kind of Young-tubes on our 800, 600 and 400 MHz NMRs.

often we are using the beige ones, ... these are out of PEEK. These spinners
are much more stable at different temperatures - like the ceramics one. Also
there is a manufacturer of these - http://www.rototec-spintec.com - these are
selling also a brass ring for this system.

You can put also this brass ring to the blue POM spinners or the Kel/F ones. This
makes the spinner a bit more heavy. Also it keeps the drop into the magnet a bit
more save, because there is the eddy current brake effect. Especially at the 800
and 600 it is a good choice to get a smoother drop into the probe.

But it makes the system a bit more heavy - so you have to adjust a harder lifting
stream for this purpose.

Yours sincerely,

Ulrich Haunz

Am Montag, Juli 06, 2026 15:00 CEST, schrieb "Justin Douglas via groups.io"=
 <justindouglas=ku.edu_at_groups.io>:

> Hello,
>
> J Young tubes are getting stuck on occasion during the loading step in our
> walkup Bruker 400 with sample case. I believe the issue is that these top-heavy
> tubes are bouncing into a groove in the top of the upper barrel, then wedging
> themselves diagonally between these grooves and the shuttle of the autosampler.
>
> I made a movie of the issue and have uploaded with annotations to youtube
> https://youtu.be/zpM9dHy0qWE
>
> The system will try to reinsert 5 times before giving up. Many times one
> of these attempts is successful. Sometimes the tube will not load after
> five times and the systems goes into an error state, which requires manual
> intervention to restart the queue. Even if the sample loads successfully on
> the 2nd, 3rd, etc. try the extra time backs up the queue and seems to irritate
> the students.
>
> I would appreciate ideas to test my hypothesis (that tubes are jammed diagonally
> into between grooves on the BST and the sample shuttle) and ideas on how to
> address this issue. I can't imagine I am the first or only person to have this issue.
>
> A couple of additional comments:
>
> These JYoung tubes were "shortened" by a glassblower to address the height
> requirements of the sample case. Hence, it is at least conceivable that
> tubes/valves have asymmetry. We have a tool to check tube straightness and
> all of the tubes are acceptable. The tool doesn't check the valve, though.
> Given that all tubes will load at least some of the time AND the amount
> of money already spent on these J Young tubes, I would anticipate skepticism
> (to put it politely) from the faculty if I suggest the issue is with the tubes.
> In addition, there is another instrument right next to this 400 and (to our
> knowledge) no JYoung tubes have ever gotten stuck in its sample case! And
> I'll also note that (to our knowledge) no standard 5mm tubes from any vendors
> (Wilmad, Norell, New Era, etc.) has ever gotten stuck in either sample case.
>
> One faculty member suggested trying to rig some sort of "bumper" that fits
> on top of the shuttle and dampens the oscillations of the JYoung tube (but
> presumably have no impact on normal tubes). Having spent the better part
> of an afternoon perched on a ladder watching tubes go in and out of the instrument,
> I think it is conceivable, but it seems like there is a lot that could go
> wrong just above the bore of the magnet, which makes me nervous.
>
> Thanks,
>
> Justin
>



--
=========================

Ulrich Haunz, Dipl.-Ing. (FH)
NMR-Spektroskopie

Raum L 710, L 807
NMR-Center M520, M521, M527
Postfach M 715
Fachbereich Chemie
Universitaet Konstanz

78457 Konstanz

Tel.: +49 (0)7531 88-4261 -> M521 Bruker Avance NEO 800
Tel.: +49 (0)7531 88-4422 -> L807 JEOL Lambda 400
Tel.: +49 (0)7531 88-4955 -> L710 Bruker AVIII HD 400
Tel.: +49 (0)7531 88-4956 -> M520a Bruker AVIII 600
Tel.: +49 (0)7531 88-4957 -> M527 Bruker AVIII 400 Automation
Tel.: +49 (0)7531 88-4958 -> M520 Bruker AVIII 400 WB
Tel.: +49 (0)170 9153213
FAX : +49 (0)7531 88-3898
=========================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3547): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3547__;!!PvDODwlR4mBZyAb0!QNg552K5sF6zFltbg0RaPp32_JQCNq=
Q7298IWZxMa-cp0CXQeByGaaORJOfLlgyJx9BZ_htTy_Giiijqy7vTku8GV18n$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120139056=
/7559972__;!!PvDODwlR4mBZyAb0!QNg552K5sF6zFltbg0RaPp32_JQCNqQ7298IWZxMa-cp0=
CXQeByGaaORJOfLlgyJx9BZ_htTy_Giiijqy7vTkl-Nj7TS$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-


Received on Tue Jul 07 2026 - 06:52:12 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 01 2026 - 17:50:19 MST