Re: [AMMRL] Helium Transfer Lines - Refreshing vacuum. #poll-notice

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Wed, 30 Jul 2025 09:26:42 -0700

Just to follow-up... The reason I ask is the engineer mentioned that most
people wait far too long.  By the time you see evidence of it having
gone soft, you have likely been losing quite a bit of helium due to excessive
boiloff for some time.

Apparently most transfer lines also come with the necessary equipment to pump
it down.  They were some of the bit and pieces lying around the lab that
I had no idea of the purpose.


-=-=-=-=-=-=-=-=-=-=-=-