Re: [AMMRL] BSMS restart in NEO console

From: Nagarajan Murali via groups.io <nmurali29=yahoo.com_at_groups.io>
Date: Fri, 25 Jul 2025 19:32:56 +0000 (UTC)

Are you running Topspin 4.5.0?  I used to run Topspin 4.0.7 and there
wasn’t any issue. But after probe repair andits firmware update I was
asked to move to 4.5.0. But, now I have to frequently reboot the console. 

Murali Rutgers

On Friday, July 25, 2025, 11:03 AM, gregory.wylie via groups.io wrote:

Hi
Man you get 2-3 weeks! I have 3 NEO's the newest 2023 has to be rebooted every week
or it will lock up. The 2 from 2018, the longest they can go is 2 weeks. We just
have our routine of rebooting them when we do LN2 fills. 
Greg 

Gregory P. Wylie, Ph.DNMR Facility Manager
Texas A&M University
Department of Chemistry #3255
580 Ross St.
College Station, TX  77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu




-=-=-=-=-=-=-=-=-=-=-=-