Re: [External Email] Re: [AMMRL] BSMS restart in NEO console

From: Juergen Schulte via groups.io <schulte=binghamton.edu_at_groups.io>
Date: Fri, 25 Jul 2025 14:35:45 -0400

All 6 fans report as “rotating”.

I’ll try power cycling on the PDU service page when it fails again
next month.

…and I’m probably gonna leave the router alone for now😊

Best Regards,

Jürgen

_______________________________________________

Jürgen Schulte, Ph. D.
NMR Specialist
Binghamton University
SN Building, RM 1024
25 Murray Hill Road
Vestal, NY 13850

Ph.: 607-777-4405
Fax: 607-777-4478
schulte_at_binghamton.edu
http://nmr.binghamton.edu/Schulte.html

> From: main_at_ammrl.groups.io On Behalf Of Shore, Jay via groups.io
> Sent: Friday, July 25, 2025 2:18 PM
> To: main_at_ammrl.groups.io
> Subject: [External Email] Re: [AMMRL] BSMS restart in NEO console

We were having a similar problem with our Avance III and it turned out to be
an overheating issue. After replacing a fan no problem...so far.

I do not know about NEO consoles, but I thought it might be something to consider.

Good luck.

Jay



  _____

From: main_at_ammrl.groups.io on behalf of Michael Groves via groups.io
> Sent: Friday, July 25, 2025 12:40 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] BSMS restart in NEO console

In the NEO it's a router and it's not as simple as just swapping out the switch
in the older models of console. It's specifically configured and the spectrometer
has its own subnet... You can update the router firmware and that sometimes helps
with weird communication issues. However if the unthinkable happens and the router
configuration gets set back to the factory default as part of the firmware update
(shouldn't happen, but it is possible) then you'll either have to replace the router,
or if you're lucky we can help reload the correct configuration.

Also note--this is for the Cisco routers in some NEO consoles. The MikroTik
routers are a whole different beast and I'm not sure how much we can do with
those. But they don't seem to have the same issues that we run into with
the Cisco routers.

Cheers,

Mike

On Friday, July 25, 2025 at 11:35:38 AM MDT, Spin-Doc <bert.heise_at_spin-doc.net wrote:

Hi,

Maybe I‘m captain obvious but have you already tried to replace the
internal ethernet switch with a different one to see if it helps with the
underlying problem?


Cheers

Bert

-> Sent via mobile <-

Spin-Doc

Magnetic resonance equipment, service, training & more

Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany

Web: http://www.spin-doc.net

Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab, Stelar & Barthel HF


Am 25.07.2025 um 17:51 schrieb Wright, Marcus via groups.io:

I assume you have tried the following:

type ha for hardware access: (File ending with 5549) You can click on the
BSMS Service GUI open button and click on the Shut Down button under BSMS status
(file ending with 5549). Or you can click on the BSMS/2 ELCB open button then
click on Service at the top of the list (file ending with 5551). Then click
on Reset ELCB (file ending with 5552)

Marcus

Marcus Wright Ph.D.
Nuclear Magnetic Resonance Laboratory
Wake Forest University


On Fri, Jul 25, 2025 at 10:22 AM Juergen Schulte via groups.io wrote:

Is there a way to selectively power-cycle an individual component in a NEO
without having to reboot the whole console?

In our case, every 2-3 weeks the BSMS loses connection to the rest of its NEO
400 console and to Topspin. The restart buttons on the BSMS service webpage
won’t restart it, and only a spectrometer reboot will get it going again.

Best Regards,

Jürgen

_______________________________________________

Jürgen Schulte, Ph. D.
NMR Specialist
Binghamton University
SN Building, RM 1024
25 Murray Hill Road
Vestal, NY 13850

Ph.: 607-777-4405
Fax: 607-777-4478
schulte_at_binghamton.edu
http://nmr.binghamton.edu/Schulte.html





-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2575): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2575__;!!PvDODwlR4mBZyAb0!XGbmKjdMw8Bjjvy4dfsXBapbPl3fs3=
WlOjfuO4KrR68d4P3ZMoMnNNL9xkLXBP4mt493DwLYHMRCUVr-_aQdWVpbsxZx$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114342729=
/7559972__;!!PvDODwlR4mBZyAb0!XGbmKjdMw8Bjjvy4dfsXBapbPl3fs3WlOjfuO4KrR68d4=
P3ZMoMnNNL9xkLXBP4mt493DwLYHMRCUVr-_aQdWarrTkfn$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jul 25 2025 - 11:49:16 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST