Re: [AMMRL] BSMS restart in NEO console

From: Wright, Marcus via groups.io <wrightmw=wfu.edu_at_groups.io>
Date: Fri, 25 Jul 2025 11:04:10 -0400

I assume you have tried the following:

type ha for hardware access: (File ending with 5549) You can click on the
BSMS Service GUI open button and click on the Shut Down button under BSMS
status (file ending with 5549). Or you can click on the BSMS/2 ELCB open
button then click on Service at the top of the list (file ending with
5551). Then click on Reset ELCB (file ending with 5552)

Marcus

Marcus Wright Ph.D.
Nuclear Magnetic Resonance Laboratory
Wake Forest University


On Fri, Jul 25, 2025 at 10:22 AM Juergen Schulte via groups.io wrote:

> Is there a way to selectively power-cycle an individual component in a NEO
> without having to reboot the whole console?
>
> In our case, every 2-3 weeks the BSMS loses connection to the rest of its
> NEO 400 console and to Topspin. The restart buttons on the BSMS service
> webpage won’t restart it, and only a spectrometer reboot will get it going
> again.
>
>
>
> Best Regards,
>
> Jürgen
>
>
> *Jürgen Schulte, Ph. D.
> *NMR Specialist
> *Binghamton University
> *SN Building, RM 1024*
> *25 Murray Hill Road*
> *Vestal, NY 13850*
>
> *Ph.: 607-777-4405*
> *Fax: 607-777-4478*
> schulte_at_binghamton.edu
> http://nmr.binghamton.edu/Schulte.html
>
>
>
>
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2569): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2569__;!!PvDODwlR4mBZyAb0!V0giXTv3_PxV5VPAewPtyQ9YlMsOBq=
87XMXh-MyOLxdEmwFphhlwNidNC-PSh6XIScBFxvLqIDP1p7y7Az_TivgIfgZJ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114338328=
/7559972__;!!PvDODwlR4mBZyAb0!V0giXTv3_PxV5VPAewPtyQ9YlMsOBq87XMXh-MyOLxdEm=
wFphhlwNidNC-PSh6XIScBFxvLqIDP1p7y7Az_Tir9R4yPb$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






1000005549.jpg
(image/jpeg attachment: 1000005549.jpg)

1000005550.jpg
(image/jpeg attachment: 1000005550.jpg)

1000005551.jpg
(image/jpeg attachment: 1000005551.jpg)

1000005552.jpg
(image/jpeg attachment: 1000005552.jpg)

Received on Fri Jul 25 2025 - 08:49:14 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST