Re: [AMMRL] BSMS restart in NEO console

From: Matthias Brandl via groups.io <matthias.brandl=tum.de_at_groups.io>
Date: Fri, 25 Jul 2025 16:35:49 +0200

Dear Jürgen,

The power distribution within the NEO consoles is handled by so-called
"power delivery units" (PDUs). Restarting an individual unit can be done
either physically by re-plugging the respective cord on the back of the
PDU (through the back door of the console cabinet), or electronically.

The PDUs typically are accessible via the spectrometer network under
149.236.99.9:85 and 149.236.99.9:86 (possibly also the subsequent ports
if there are more than two, probably only the former if there is only one).

Username is "nmrsu", Password is "topspinpdu", at least in our machines.
 From within the webinterface, one can switch individual outputs (there
also is a one-click "power cycle" option!). Sometimes they are even
already labelled with the connected unit; if not, just trace the cords
in the back of the console.

Best,
Matthias


On 25/07/2025 16:18, Juergen Schulte via groups.io wrote:
> Is there a way to selectively power-cycle an individual component in a
> NEO without having to reboot the whole console?
>
> In our case, every 2-3 weeks the BSMS loses connection to the rest of
> its NEO 400 console and to Topspin. The restart buttons on the BSMS
> service webpage won’t restart it, and only a spectrometer reboot will
> get it going again.
>
> *Best Regards,*
>
> *Jürgen***
>
> *_______________________________________________***
>
> *Jürgen Schulte, Ph. D. *
>
> *NMR Specialist *
>
> *Binghamton University *
>
> *SN Building, RM 1024*
>
> *25 Murray Hill Road*
>
> *Vestal, NY 13850*
>
> *Ph.:   607-777-4405*
>
> *Fax:  607-777-4478*
>
> *schulte_at_binghamton.edu* <mailto:schulte_at_binghamton.edu>**
>
> *http://nmr.binghamton.edu/Schulte.html* <http://nmr.binghamton.edu/
> Schulte.html>**
>
>

--
Dr. rer. nat. Matthias Brandl (he/him or they/them)

Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)

Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen

Phone +49 89 289-52617

matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2567): https://ammrl.groups.io/g/main/message/2567
Mute This Topic: https://groups.io/mt/114338328/7559972
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/xyzzy [walt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-


Received on Fri Jul 25 2025 - 08:02:50 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST