Hi Andre,

Are you sure that the 1H channel was really tuned down to 19F at NEO spectrometer
and not a BB channel?

For the relatively modern NMR probes with the “smart probe”
like 1H resonator (BBO probes), the tuning range for 1H is quite restricted.

The correct routing/preamplifier band can be tested by 50 ohm RF load (input/output
of any 50 ohm attenuator) connected instead of the probe. Wobb curve should jump
to zero entirely.

In any case, running ATMM is a very good suggestion to try out.

The older TXI probes had the atma system without position tracking, so no
“memory” about any positions, but I don’t think
this is your case

Regards,

Dr Pavel Iouchmanov,

P&L Scientific AB
Repair/customization of NMR probes and spectrometers
Electrophoretic NMR and custom NMR solution for electrochemistry

+46 736 220 672
Applied Physical Chemistry,KTH
Teknikringen 36,
100 44 Stockholm
Sweden
pavel_at_plscientific.se ( pavel_at_plscientific.se )
http://www.plscientific.se


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2492): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2492__;!!PvDODwlR4mBZyAb0!Qe4Q8CRbSBIoIXQK0lvhUJWNAcXdWs=
Izolsx20YA0M9uhPDuJjj0HqG7iSCh7nWJa1hiK2gJosGglMEk-Hc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114154407=
/7559972__;!!PvDODwlR4mBZyAb0!Qe4Q8CRbSBIoIXQK0lvhUJWNAcXdWsIzolsx20YA0M9uh=
PDuJjj0HqG7iSCh7nWJa1hiK2gJosGgNiyzFMM$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Jul 15 2025 - 13:51:51 MST

This archive was generated by hypermail 2.4.0 : Sat Aug 02 2025 - 14:13:11 MST