Hi Andre,
I have a feeling that the 1H HPPR on your AV I is not 19F capable. Do
you want to reach 19F while the 1H channel is connected to the 1H HPLNA
or do you use the XBB19F module to run 19F?
On a NEO system the 1H/19F HPLNA is capable of doing both nuclei.
Clemens
On 7/14/2025 4:29 AM, Andre Dallmann via groups.io wrote:
>
> Dear all,
>
> I was wondering, whether anyone has ever tried to change the starting
> values in the .bb for 19F/1H. I can see all the heteronuclei, but I want
> to detune the 1H channel to 19F.
>
> I know it is possible, since on our NEO400 it worked, however on the AVI
> (where the probe head is supposed to be installed) the atma cannot not
> quite get to the point so another starting point closer to 19F would be
> great.
>
> For the heteronuclei I had similar problems and shifting the starting
> values did the trick.
>
> Best, André
>
> --
> Dr. André Dallmann
> Leiter der NMR-Abteilung
>
> Email: andre.dallmann_at_chemie.hu-berlin.de
> Raum: 0'103
> Telefon: 82952
> Handy: 01797657855
> ORCID-ID: https://orcid.org/0000-0002-0141-2635
> Webseite:
> https://www.chemie.hu-berlin.de/en/forschung-en/nmr-spectroscopy/nmr-spectroscopy?set_language=en
>
> Adresse:
> Humboldt Universität zu Berlin
> Institut für Chemie
> AG NMR
> Brook-Taylor-Str. 2
> 12489 Berlin
> Deutschland
>
>
>
>
>
>
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2496): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2496__;!!PvDODwlR4mBZyAb0!QfxBdpkJOqlKwyyr-xj06E-4DIFWMw=
hNYpKLqyhIQu-MQG-Z6bdNKy7k0Xn0C1jh7-suaKYvobgTH14wAfB8aWQRMwbw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114154407=
/7559972__;!!PvDODwlR4mBZyAb0!QfxBdpkJOqlKwyyr-xj06E-4DIFWMwhNYpKLqyhIQu-MQ=
G-Z6bdNKy7k0Xn0C1jh7-suaKYvobgTH14wAfB8aYVfeG51$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Jul 15 2025 - 14:13:06 MST