Karel,
You have a point but I do not think you have to set it for all solvents.
You can either set a global shimroutine in the ICONNMR configurator in Lock/Shim
Options -> Shim Program or you can go to the submenu Solvent/Probe Dependencies
and set individual shim routines for different solvents. And for individual
choices the options have to be in the shim method selector and the user has
to be give the right to edit lock and shim method.
Outside of ICONNMR I always work with "to pshim gui" so I can set the options.
Clemens
On 1/20/2026 8:59 AM, Karel Klika via groups.io wrote:
>
> Hi Clemens,
>
> This should certainly work, but you would have to set it for all solvents?
> If this is what you want that is fine, but if not then I guess a cumbersome
> work around would be to have double entries for the each of the solvents you
> want it for, one for just normal shimming and the other for the other case
> and the user can choose in icon which procedure s/he wants by choosing one
> of the two solvents.
>
> A second thing is that if others users change the shim selection in the BSMS
> then this is then activated. If the instrument is only in icon use this would
> be fine but maybe not if it was in hybrid use.
>
> I tried to implement the preference by running a macro in the automation
> acquisition au program (xcmd ("macro fielename") but it would not run (the
> au program would compile though and the macro from the command line ran as
> desired). I am using TS3.7 running under Windows 10 on AVIII. Anything to
> do with Defender ?
>
> Best regards,
> Karel
> ________________________________________
> From: main_at_ammrl.groups.io on behalf of Anklin, Clemens via groups.io
> Sent: 19 January 2026 22:22:12
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Topshim with parameters and Autoshim turned on
>
> Hi Andre,
>
> set the shim program in the ICONNMR configurator to
> "TOPSHIM_WITH_AUTOSHIM convomp tuneay tuneax"
>
> this should do the job
>
> the shims to use during autoshim are as usual activated from the BSMS
> panel.
>
> Best regards
> Clemens
>
>
> On 1/19/2026 12:56 PM, Andre Dallmann via groups.io wrote:
>>
>> Dear all,
>>
>> has anyone a quick way to get ICON to make "topshim convcomp tunay
>> tuneax" and afterwafrds turning on AUTOSHIM (ideally with z1z2xy)?
>>
>> Sure this must be easy, but somehow cannot find the solution...
>>
>> Thanks and best, André
>>
>> --
>> Dr. André Dallmann
>> Leiter der NMR-Abteilung
>>
>> Email: andre.dallmann_at_chemie.hu-berlin.de
>> Raum: 0'103
>> Telefon: 82952
>> Handy: 01797657855
>> ORCID-ID: https://orcid.org/0000-0002-0141-2635
>> Webseite:
>> https://www.chemie.hu-berlin.de/en/forschung-en/nmr-spectroscopy/nmr-spectroscopy?set_language=en
>>
>> Adresse:
>> Humboldt Universität zu Berlin
>> Institut für Chemie
>> AG NMR
>> Brook-Taylor-Str. 2
>> 12489 Berlin
>> Deutschland
>>
>>
> --
> =========================
==========================
=====================
> Clemens Anklin Ph.D. Vice President
> Bruker BioSpin NMR Applications & Training
> 15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
> Billerica MA 01821 bluesky: _at_canklin.bsky.social
> web: http://www.bruker.com
> e-mail: clemens.anklin_at_bruker.com
> =========================
==========================
=====================
>
>
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3136): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3136__;!!PvDODwlR4mBZyAb0!Xt1vZvkU0kBO9z-3PwcDH-RVjOsSC1=
fRgONSnch-uijjnD0Bxi9ZHai3n1Eh891YigXLYIcYrqqz9a17VdZwD3xadHo5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117349140=
/7559972__;!!PvDODwlR4mBZyAb0!Xt1vZvkU0kBO9z-3PwcDH-RVjOsSC1fRgONSnch-uijjn=
D0Bxi9ZHai3n1Eh891YigXLYIcYrqqz9a17VdZwDykDTqiW$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Jan 20 2026 - 07:04:06 MST