Hi TJ,

Bruce Adams used inversion recovery for this sort of application, typically
for phosphorus. I’ve attached a document describing the approach as
well as a couple of pulse sequences for 1) single time observation and 2)
pseudo-2D time series measurement. Hope it helps!

Best,
Walt

-------------------------------

Walter Massefski, Ph.D.
Director
Department of Chemistry Instrumentation Facility
Massachusetts Institute of Technology
77 Mass. Ave. Rm: 18-0090
Cambridge, MA 02139
T: (617) 253-2016
wwm_at_mit.edu <mailto:wwm_at_mit.edu>

> On Jan 6, 2026, at 3:57 PM, Thusitha Jayasundera via groups.io wrote:
>
> Hello all--I am trying to adopt the sequence followed in the attached article
> (including the supplementary material) that studied a kinetic process via
> proton NMR to a system observing 31P instead, with no success. I was hoping
> that ideally, someone may already have a working sequence in their library
> and willing to share it.
>
> For context, the system in question has two 31P nuclei that (potentially)
> exchange, and the hope is to invert one signal with a soft 180-pulse and
> track the intensity of the second, arraying the "mixing time" between the
> 180 and final 90-detection pulse.
>
> Any advice would be much appreciated as always!
>0
> T.
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3106): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3106__;!!PvDODwlR4mBZyAb0!VDFHH-xYG2u0Ig7RkC-D5zzuTuouy4=
ByyH0tMgvAHD9Uk4OA491LstsOOtVMkPJbcZsQpah823AcapsSTOt_F0qXwFFh$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117112998=
/7559972__;!!PvDODwlR4mBZyAb0!VDFHH-xYG2u0Ig7RkC-D5zzuTuouy4ByyH0tMgvAHD9Uk=
4OA491LstsOOtVMkPJbcZsQpah823AcapsSTOt_F2PgGV-A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jan 07 2026 - 07:22:20 MST

This archive was generated by hypermail 2.4.0 : Mon Feb 09 2026 - 16:54:28 MST