Re: [AMMRL] AV2+/3: cf stuck detecting last HPPR module

From: Martha Morton via groups.io <marthadmorton=gmail.com_at_groups.io>
Date: Wed, 29 Jan 2025 08:09:34 -0600

Hi Bert,

Is there power to the cover module of the HPPR? Is there a fuse in the HPPR
that is blown?

Best,
Martha Morton


On Wed, Jan 29, 2025 at 5:13 AM Spin-Doc via groups.io wrote:

> Hi,
>
> Has anyone come across this yet:
> On an AV2+/3 mini (with IPSO), “cf” gets stuck detecting the last HPPR
> module - apparently irrespective of how many preamp modules are plugged in,
> see screenshot.
> The setup in the screenshot has one dual 1H/2H module (P1/P2) and an X
> channel module - cf ist stuck at P3. If we plug in another module above or
> below then it gets stuck at P4. Tried swapping the order if preamps
> shuffling them around but it made no difference. Same behaviour with
> another top/cover unit.
> Any ideas…?
>
> Best regards
> Bert
>
>
>
>
> Spin-Doc
> Magnetic resonance equipment, service, training & more
> Dr. rer. nat. Bert Heise
> Owner, applications scientist
> Kiefernweg 13
> 58239 Schwerte
> Germany
> Web: http://www.spin-doc.net
> Mail: bert.heise_at_spin-doc.net
> Phone: +49-171-5135315
>
> Representing MestreLab, Stelar & Barthel HF
>
>

--
Martha Morton, PhD
Director of Research Instrumentation, University of Nebraska Lincoln
Safety Co-Chair, Department of Chemistry
Nebraska ACS, Chair 2021
ACS, MWRM Chair 2021


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1987): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1987__;!!PvDODwlR4mBZyAb0!Rs-xIzPHz3t_iT6nNtV6civ2EA-SMG=
9Y61ztRLu1TIEFuy048xvpvRM52mGKs5MrcxRPELIt8lFZDEcAqYuoHPwy3IQM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/110876143=
/7559972__;!!PvDODwlR4mBZyAb0!Rs-xIzPHz3t_iT6nNtV6civ2EA-SMG9Y61ztRLu1TIEFu=
y048xvpvRM52mGKs5MrcxRPELIt8lFZDEcAqYuoHII_lHqf$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jan 29 2025 - 06:09:54 MST

This archive was generated by hypermail 2.4.0 : Sat Feb 01 2025 - 17:12:42 MST