Re: [AMMRL] Helium push gas assembly?

From: gregory.wylie via groups.io <gregory.wylie=chem.tamu.edu_at_groups.io>
Date: Mon, 27 Jan 2025 07:50:02 -0800

Hi Ken

First let me say we own our dewars since we have helium recovery and we wanted
a quick way to connect to the dewar vent port. We do a lot of transfers from
the liquifier to the dewars and then from dewars to magnets, at least one of
these events a week if not 2.

We put a 1/2 inch NPT nipple in the 1/2 inch ball valve of the main vent and
put a 1/2 inch NPT to KF-25 adapter on it (pic Dewar_vent_port). This allows
us to quickly attach the dewars to the helium recovery system for filling or
to add the valve assembly for doing magnet fills.

The valve assembly (pic valve_assembly) is all 1/2 inch NPT fittings and ball
valves, a KF-25 adapter and a push-to-connect fitting for 1/4 inch tube. We
use the old style rotameter and 0-3 PSI gauge assembly that many people are
familiar with (pic flow_assembly). This connects to the high pressure helium
tank like normal via 1/4 inch push-to-connect, and the out flow is also 1/4
inch push-to-connect that then connects to the valve assembly. The valve assembly
lets us close off this flow, vent the dewar. There is a pic (In_operation) of
everything on a dewar. When we purchase/rent a dewar for fills we do add a
nipple and KF-25 adapter to the vent port of these dewars while in our
possession and remove it when it is returned.

Greg

Gregory P. Wylie, Ph.D
NMR Facility Manager
Texas A&M University
Department of Chemistry
P.O. Box 30012
College Station, TX  77843-3255
gpwylie_at_tamu.edu
979.458.0705 (voice)
979.845.4719 (fax)
706.206.0007 (cell)
http://nmr.tamu.edu


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1979): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1979__;!!PvDODwlR4mBZyAb0!UgXnyMoNqFqI_sjS7whvVbNWnBxCsZ=
yl6EDB3H_a7KPaM4h47-0dpR4pL0gXtxwqZncOjwYQ6BLQJ-VL_glu4iBeciZW$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/110838638=
/7559972__;!!PvDODwlR4mBZyAb0!UgXnyMoNqFqI_sjS7whvVbNWnBxCsZyl6EDB3H_a7KPaM=
4h47-0dpR4pL0gXtxwqZncOjwYQ6BLQJ-VL_glu4qmgdrrQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






Flow_assembly.jpg
(image/jpeg attachment: Flow_assembly.jpg)

In_operation.jpg
(image/jpeg attachment: In_operation.jpg)

Dewar_vent_port.jpg
(image/jpeg attachment: Dewar_vent_port.jpg)

Valve_assembly.jpg
(image/jpeg attachment: Valve_assembly.jpg)

Received on Mon Jan 27 2025 - 08:04:02 MST

This archive was generated by hypermail 2.4.0 : Sat Feb 01 2025 - 17:12:42 MST