Maybe they’ve changed the way they set up the QT NMR systems? (ours
was installed in 2021)

In our case (7 magnets 300-500 plus a pumped 800, and a 200L dewar that we
use for fills, which almost always has some liquid in it) the large compressor
only goes on when doing transfers or fills. The rest of the time a small pump
in the liquefier controls the feed of gas from the header- and we have never
had an issue with it pulling excess boiloff from the magnets. A regulator
controls the feed in from the storage tanks- adding to what comes from the header.

Our liquefier runs 24/7.

And yes, the learning curve was kind of steep and annoying 😊 but
we’ve pretty much got it all sorted out now…

Tara

*********************************************
Dr. Tara Sprules

0.6 FTE, Working days: Tuesday, Wednesday and Friday
QANUC
Quebec/Eastern Canada High Field NMR Facility

http://www.nmrlab.mcgill.ca

phone:  (514) 398-1721
fax:    (514) 398-8254

3420 University St., Rm 023
McGill University
Montreal, QC, H3A 2A7

SHIPPING ADDRESS:

Department of Chemistry, McGill University

Rm. 41 801 Sherbrooke St. W

Montreal QC

H3A 0B8

*********************************************

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#962): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/962__;!!PvDODwlR4mBZyAb0!VxABoEsQg5uMcwpLIbXrOjMJpJUt8vuC=
3RruZf0Qn2HzGTonpjZJTAoxnOZ47GqnCT9JthzC1mkU21RQOStHyR0$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104018806=
/7559972__;!!PvDODwlR4mBZyAb0!VxABoEsQg5uMcwpLIbXrOjMJpJUt8vuC3RruZf0Qn2HzG=
TonpjZJTAoxnOZ47GqnCT9JthzC1mkU21RQxf_GQHI$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Jan 31 2024 - 08:40:48 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 01 2024 - 14:56:24 MST