Re: [AMMRL] BMPC2 email/internet comm

From: Platz Hirsch <ammrl_at_haessner.com>
Date: Thu, 18 Jan 2024 16:14:34 +0000

Dear Mike,

a few years ago during my active work in Munich I created a complete solution
to connect the BMBC computer via an additional gateway computer to the internet.

This gave you access to a bunch of values and made a complete remote control
of this unit using the RDP protocl possible.

This was possible without security issues, because the gateway computer excluded
any direct routing. You could apply any kind of security configuration to the
gateway computer.

Data became extracted from the BMPC - I belive they internally use an access
database - converted and stored within a munin database and at least during
my active time displayed using the rrdtools.

Some alerts were configured too.

As far as I know this installation is still running.

That time I made a step by step explanation, how to install everything. It's
written in german language. I am rather sure, that this tutorial still exists.

I have no access and of course cannot decide about this material. But the
people, now responsible, are reading here and they will get in touch with
you with some probability.

Greetings

Rainer

Von: main_at_ammrl.groups.io <main_at_ammrl.groups.io> Im Auftrag von Bob Berno
Gesendet: Donnerstag, 18. Januar 2024 16:39
An: main_at_ammrl.groups.io
Betreff: Re: [AMMRL] BMPC2 email/internet comm

Hello Mike,

I can't answer your question since I still have the older model BMPC2 with
an older version of Windows which is NOT on the network.
But I wanted to ask you a question about your newer computer system.
Does it still use the analog phone line to send you text message alerts?

I agree that it would be nice to be able to get regular email status updates
from the BMPC/2 instead of having to stand and look at the workstation monitor.
But I would not want to rely only on internet/network messages.

Cheers,
Bob.
On 2024-01-18 9:39 a.m., Delk, Michael Joseph via groups.io wrote:

Hello All,

We have had a new BMPC2 installed on our older 800 pumped Bruker magnet recently.
This comes with a newer Siemens industrial type PC running Windows10. We have
a successful internet connection allowing remote login as well internet browser etc.

So far, we have nit successfully got an email sent out to me from the unit
software. Also, would like to get the alarm notifications working in this
manner somehow.

Interested to hear back on ideas of successful modern BMPC2 comm solutions.

Regards,
Mike Delk
NMR Manager
University of Pittsburgh School of Medicine
Department of Structural Biology
412.370.0066--


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#942): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/942__;!!PvDODwlR4mBZyAb0!S3w_bUGay1kqPl-uddI4Sswg-qBB0lyG=
jZgod1rXb-E_Olw4YQd8CJLjJ5lmrTHk4QJ3RRW4bKfOpg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/103809147=
/7559972__;!!PvDODwlR4mBZyAb0!S3w_bUGay1kqPl-uddI4Sswg-qBB0lyGjZgod1rXb-E_O=
lw4YQd8CJLjJ5lmrTHk4QJ3RRVH5aHMGQ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Jan 18 2024 - 08:32:50 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 01 2024 - 14:56:24 MST