[AMMRL] [SOLVED] Shimming issues

From: Kenneth Sharp-Knott <kknott_at_vt.edu>
Date: Fri, 12 Jan 2024 13:44:28 -0500

*Topline: Bruker was right from the beginning. The issue was a problem
with the VT gas flow stemming from a malfunction in the BVT3000.*

After replacing every inch of tubing to the console and from the VT unit to
the probe trying to find a leak, blockage or kink, I noticed that the float
ball in my BVT 3000 did not drop when it was turned off. Fortunately, I
have lots of old instruments with BVT3000's and swapped them out.

Turns out the old unit was obstructing the flow in some way. The flow did
increase/decrease, but was evidently less than 100%. I was unable to tell
a significant difference but clearly, there was one. First clue after
replacement was that the power required to regulate at 25C went from 1.1%
to 0.5%.

After replacing the BVT 3000 and redoing 3D shims I was able to get
excellent lineshape and my D2O samples look excellent. As to the Co59
sample, it still looks a little misshapen, but I'm getting 50Hz instead of
3-4ppm. Not perfect, but likely better than it's been in a long time.

I should have probably caught this sooner, but the BVT3000 seemed to be
regulating the flow.

On another note, I think it would be nice to get a flow meter from these
kinds of flow rates. Any suggestions? The only ones I have found are for
much lower flow rates.

Thank you all for eventually getting me pointed in the right direction.
I've learned a ton about VT, temperature gradients, convection, Co59 and
many other things that I will employ in the future!


--
Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#931): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/931__;!!PvDODwlR4mBZyAb0!Xi4j2tJ8Rbs7N1_wNr2KBuMfvNw8noWj=
_OxOjvor5AKqYvjGT5kU_cuBQaL52qV3sRgvSwinxXvPLA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/103689222=
/7559972__;!!PvDODwlR4mBZyAb0!Xi4j2tJ8Rbs7N1_wNr2KBuMfvNw8noWj_OxOjvor5AKqY=
vjGT5kU_cuBQaL52qV3sRgvSwjrOeSIcw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jan 12 2024 - 10:48:39 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 01 2024 - 14:56:24 MST