Re: [AMMRL] DMSO-D2O sample lock issue

From: DrSpin <rajanp_at_uw.edu>
Date: Fri, 12 Jan 2024 10:20:40 -0800

Hi Tracy
    If you are talking about issues with mixture of solvents that have more
than one 2H NMR resonance, at least the Topspin 3.X onwards (i.e.
instruments that are AVance-III or newer) has an elegant way to deal with
it via the 'edlock' interface. I am of course assuming that your DMSO is
actually DMSO-d6. You can create a new solvent entry in the table and
indicate the actual resonance positions (4.7 ppm for D2O, etc.) and then
specify which resonance peak that will be used for Locking.
   That's my 2c and I am sure wiser heads are going to weigh in with better
answers as I write this !

Best Regards
Rajan

On Fri, Jan 12, 2024 at 10:07 AM Zhang, Tracy Chunxin (GE Aerospace, US) wrote:

> Dear all,
>
>
> I have run into an odd issue here: one of the chemists using my walk-up
> instrument, a 500MHz Bruker Avance III, said she had trouble locking her
> sample in a DMSO-D2O mixture (80/20) in ICON. When I tried to run it
> manually, you can see the lock signal (we used to call them “rabbit ears”
> in grad. School) sweeping in the lockdisp window when you first dropped the
> sample in the magnet, but a few seconds later, the lock signal turns into
> flat noise and the sample would not lock. A D2O, DMSO, and DMSO/D2O
> (80/20) standards all worked fine. The sample worked fine on our Agilent
> 600MHz NMRS. The chemist could not think of anything in her sample that
> might have possibly caused the problem. Just wonder if any of you have
> seen something similar and if there is a work around. We would prefer for
> her to be able to run her own samples on the walk-up instrument. Thank
> you!
>
>
>
> PS—Wish I could attach the short video of the disappearing lock signal,
> but our firewall blocks any external email attachments. ☹
>
>
>
> *Tracy Zhang, PhD*
>
> Analytical Chemist
>
> GE Aerospace
>
>
>
> T +1 518 387 7642
>
>
>
>

--

____________________________________

Rajan K Paranji, Ph.D.NMR Services Manager
*Department of Chemistry**Room 65, Bagley Hall*
University of WashingtonSeattle, WA 98195
phone : 206 685 2581 fax: 206 685 8665email: rajanp_at_uw.edu
____________________________________________________________________


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#928): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/928__;!!PvDODwlR4mBZyAb0!RV6VfV6lXOtTPFehwTaoc3c9g4CqKej3=
vVJ9i-T67VPy1yOHEjwgAv-q_6PBMKP7GpJvC5R4gyynaw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/103688379=
/7559972__;!!PvDODwlR4mBZyAb0!RV6VfV6lXOtTPFehwTaoc3c9g4CqKej3vVJ9i-T67VPy1=
yOHEjwgAv-q_6PBMKP7GpJvC5TLqRsZ-w$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Jan 12 2024 - 10:21:06 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 01 2024 - 14:56:24 MST