Re: [AMMRL] Shimming issues related to oscillating lock level

From: Mike Lumsden <Mike.Lumsden_at_dal.ca>
Date: Wed, 10 Jan 2024 02:31:31 -0800

Ken - a friendly reminder that if you are indeed battling a temperature gradient,
topshim now gives us the “tcorr” option to apply a correction.
We use tcorr routinely here for aqueous samples on a TCI Prodigy.

Even if your version of topshim does not allow “tcorr” yet,
you could achieve the same effect via a simple AU program as part of your
shimming routine in iconnmr for aqueous solvents.

Good Luck!
Mike


-=-=-=-=-=-=-=-=-=-=-=-