[AMMRL] AMMRL: turning off tube identification in SampleJet

From: Karel Klika <karkli_at_utu.fi>
Date: Sat, 6 Jan 2024 19:57:48 +0000

Dear all,

To change tube identification, which also identifies the tube diameter:

1. Run ha (either enter ha from the command line or select it from the drop down menu)
2. Select SampleJet
3. Login
4. Go to Basic service
5. Go to User settings
6. There are separate sections for racks and for the openshop-ring
7. For the option check sample diameter with matrix code, make the appropriate
        selection from the options available
8. Save settings
9. All has been changed and ready to run as desired

Regards and thanks to Walt Massefski and Thomas Smith,

Karel Klika


-=-=-=-=-=-=-=-=-=-=-=-