Dear Karel,
There is an entry for ‘Check sample diameter with matrix code’
for each of the OpenShop and the Sample Racks - it is located in the Basic Service —> User Settings page.
You can set one or both to ‘do not check’, although this will
require that you insist on a single diameter tube, since the system will not
recognize when it needs to change shuttles. We have the ‘Open Shop’ set to
‘do not check’ and the racks set to check, to give our users the flexibility
to use 3mm or 1.7mm tubes in the inner racks, but not in the outer rack.
Best,
Walt
-------------------------------
Walter Massefski, Ph.D.
Director
Department of Chemistry Instrumentation Facility
Massachusetts Institute of Technology
77 Mass. Ave. Rm: 18-0090
Cambridge, MA 02139
T: (617) 253-2016
wwm_at_mit.edu
> On Jan 5, 2024, at 12:26 PM, Karel Klika via groups.io <karkli=utu.fi_at_groups.io> wrote:
>
> Dear all,
> A couple of questions:
> 1. Is there any way to suspend the tube reading function in SampleJet?
> The reason I ask is since we do not need the tubes to be read, people will
> then be able to write on top of the cap if the function is suspended.
> 2. How often do Prodigy probes need to be sent away for service ?
> Many thanks in advance, and to cut down on the number of reply emails to
> everyone, please respond only to me and I will post a summary if appropriate.
> Best for the New Year,
> Karel Klika
>
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-