Re: [AMMRL] Topspin behind a restricted ethernet port

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Tue, 17 Feb 2026 16:55:27 +0000

Dear Alvaro

the only Bruker NMR spectrometer that periodically contacts a server outside
is the Avance Core series. All other spectrometers do not need an outside
connection.

License activation can be done online and off-line. The offline activation
involves a sneaker net connection to a computer with outside access. You
move a file to a connected computer and later move a file back to the
spectrometer

This is described in the CodeMeter manual that is part of Topspin.

With best regards

Clemens

On 2/17/2026 11:14 AM, Alvaro Herrera via groups.io wrote:

Dear colleagues,

I am wondering if anyone has experience running TopSpin on a computer running
Linux, controlling a Neo console where the computer’s network access
(Wall Port) is restricted by IT to a defined set of allowed IP addresses.

My specific concern is TopSpin’s license verification:

Does TopSpin periodically check its license online after initial activation,
or is the license validation handled entirely locally (e.g., via CodeMeter
or a local license server)?

If I upgrade to a newer TopSpin version on the same instrument computer, will
it require access to any external license or activation server, and therefore
need additional IP addresses opened by IT?

Any details about how your site has configured this would be very helpful.

Thanks,


Alvaro Herrera Ph.D.
NMR Center Director
Chemistry & Biochemistry Department
Auburn University



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3232): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3232__;!!PvDODwlR4mBZyAb0!RVxi4fVCevKI48mlQUKOFoG99qYe-o=
YSj0_qq0RhLGzDSNBerhCDTw9lzYbniYIUFgMlWbxvlDwh0gH-e5XjctaPJrMG$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/117859746=
/7559972__;!!PvDODwlR4mBZyAb0!RVxi4fVCevKI48mlQUKOFoG99qYe-oYSj0_qq0RhLGzDS=
NBerhCDTw9lzYbniYIUFgMlWbxvlDwh0gH-e5XjcoX5fpWP$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Feb 17 2026 - 09:42:46 MST