Re: [AMMRL] Bruker communication / Syntax error from AQS3 (AV2+)

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 27 Feb 2025 16:25:51 +0000 (UTC)

Hi Bert,

I'm still not sure what version console you've got, but it sounds like it might
be newer than I thought originally.  Can you run "UniTool hppr" and talk
to the HPPR?  This should come up with default address 80 and should let
you see the preamps and cover module etc.  I'm not sure if it will work
since you had a failed cf--fingers crossed that it will. =

I'd have a look at what it comes up with for the bis file from each preamp --maybe
if the bottom one has something weird in it that could point to an issue.  And
if you can get into that you can also try the autodownload function if you want to
try downloading the firmware.  

What version console is this?  Your initial email made it sound like a
pretty old console but if you're running topspin 3.7 then it's at least gotta
have an IPSO, right?  

In older versions of topspin you had to specify which tty port the HPPR was
at but in the newer versions it just sets it automatically. 
And if it's still having issues you might check the cables between the console
and preamp just in case you've got a flaky one, and you might try a swapping
the board in the back of the AQS rack that communicates to the HPPR-- I forget
which that is, but it should be fairly obvious. 

Cheers,Mike


On Thursday, February 27, 2025 at 07:24:19 AM MST, Delk, Michael Joseph via groups.io wrote:

  Hi Bert,

Have you checked the power supplies and the ACB board on back of AQS chassis?

Mike

Mike Delk
University of Pittsburgh 
School of Medicine
Department of Structural Biology
412.370.0066


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2076): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2076__;!!PvDODwlR4mBZyAb0!W356kVRPl82eM7-In_iSbPwkhjrEJ7=
G6iueavi5ZiNQnSdBQl6o_a84UmxldSwiXVWMCp-j5wAhQsocl8u8M_Cdzl3jF$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111401280=
/7559972__;!!PvDODwlR4mBZyAb0!W356kVRPl82eM7-In_iSbPwkhjrEJ7G6iueavi5ZiNQnS=
dBQl6o_a84UmxldSwiXVWMCp-j5wAhQsocl8u8M_LL734IM$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Feb 27 2025 - 08:38:03 MST

This archive was generated by hypermail 2.4.0 : Fri Feb 28 2025 - 15:51:13 MST