Hello, Ulli,

We have metal walls (steel sheeting) in one of our NMR Labs, and they have
proved to be no problem at all. We further have steel frames for our ceiling
tiles; again, no problems.

You have not mentioned where your 5-gauss lines are located, but in shielded
magnets, any chance of a problem is greatly reduced.

Dean

***********************************
Dean L. Olson, Ph.D.
Director, NMR/EPR Lab
146 Roger Adams Lab
School of Chemical Sciences
U. of Illinois at Urbana-Champaign
600 S. Mathews, Box 31-1, MC-712
Urbana, IL 61801 USA
Lab: 217-244-0564; Cell: 217-722-9432
dolson_at_illinois.edu<mailto:dolson_at_illinois.edu>
http://scs.illinois.edu/nmr
http://scs.illinois.edu/epr/
***********************************

> From: main_at_ammrl.groups.io On Behalf Of Ulrike Werner-Zwanziger via groups.io
> Sent: Monday, February 24, 2025 8:58 AM
> To: main_at_ammrl.groups.io
> Cc: Josef Zwanziger <Josef.Zwanziger_at_dal.ca>; Mike Lumsden <mike.lumsden_at_dal.ca>
> Subject: [AMMRL] AMMRL: building of structural steel

Dear Colleagues,

Dalhousie is planning a new building, which will also house the NMR facility.
We just learned that a changed plan is to build the new building using
structural steel (not concrete), and the question was raised, whether this
will be a problem for the NMR operation.

While this may not be ideal, we likely have no choice in this matter, but I
would appreciate any experiences or advice you can give me. (I couldn't find
anything in the archives). The magnets can be moved sufficiently inside the
room so that their 5 Gauss lines can be kept away from the walls. My feeling
is as long as we stay sufficiently far away, we should be fine. Do you agree?

Happy resonating,

Ulli
--
Dr. Ulrike Werner-Zwanziger
Adjunct Professor
Department of Chemistry, NMR-3,
Dalhousie University
6245 Castine Way (!!!new address!!!)
PO BOX 15000
Halifax, NS B3H 4R2

e-mail: Ulli.Zwanziger_at_Dal.ca<mailto:Ulli.Zwanziger_at_Dal.ca>
Tel. 1 902 494 8085
FAX: 1 902 494 1310



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2060): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2060__;!!PvDODwlR4mBZyAb0!TcTn17OrzXrT1uMkhDH1Q_l0Gp8Jhj=
va2JPOUvCUJWhFv5p0k2II4Q25CAhXxiVECdkV-WTVq2p-vJT18GmJy1ZiozBy$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111358893=
/7559972__;!!PvDODwlR4mBZyAb0!TcTn17OrzXrT1uMkhDH1Q_l0Gp8Jhjva2JPOUvCUJWhFv=
5p0k2II4Q25CAhXxiVECdkV-WTVq2p-vJT18GmJyzXV7D02$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Feb 24 2025 - 07:55:19 MST

This archive was generated by hypermail 2.4.0 : Fri Feb 28 2025 - 15:51:13 MST