Hi Scott,
We have had some success by running the appropriate version of Linux inside
Oracle’s Virtual Box. I.e. set up a virtual version of Linux running
inside a modern Linux. Both Topspin and Paravision work and can collect data.
I caution that it is a bit clunky. Also it was beyond us to do. We had to have
someone who really knew what they were doing set it up.
Hope this helps!
Jeff
On Feb 14, 2025, at 10:26 AM, Scott Van Bramer via groups.io wrote:
Since Centos 7 has gone end of life we are trying to migrate Topspin 2.1 to a
new Linux system. Looking for anyone who has gotten this to work. Our spectrometer
is too old to migrate to a newer version of Topspin.
Thanks,
Scott Van Bramer
--
Scott Van Bramer
Professor of Chemistry
Director of Foundations and Transformations Program
Widener University
One University Place, Chester PA 19013
610-499-4516
sevanbramer_at_widener.edu<mailto:sevanbramer_at_widener.edu>
Zoom Office: https://widener.zoom.us/my/sevanbramer
-=-=-=-=-=-=-=-=-=-=-=-