I haven't actually done this, and making the ancillary bits work might take
a bit of trial and error (Rajan mentioned NFS for example). You will almost
certainly need to install 32 bit libraries.
The main problem, however, is that the versions of diskless and flexLM in
TS 2.1 are way too old. From notes I have, the versions in TS 3.2pl7 will
work with CentOS 7. I'm fairly sure I've heard similarly about versions in
TS 3.5pl7. So you start with the newer Topspin that can't run your old
instrument, do a custom install of diskless and flexlm, then do a custom
install of TS 2.1 and deselect those packages.
Good luck,
Andrew
On Fri, Feb 14, 2025 at 1:36 PM Scott Van Bramer via groups.io wrote:
> Since Centos 7 has gone end of life we are trying to migrate Topspin 2.1
> to a new Linux system. Looking for anyone who has gotten this to work. Our
> spectrometer is too old to migrate to a newer version of Topspin. Thanks,
> Scott Van Bramer
>
> --
> Scott Van Bramer
> Professor of Chemistry
> Director of Foundations and Transformations Program
> Widener University
> One University Place, Chester PA 19013
> 610-499-4516
> sevanbramer_at_widener.edu
>
> Zoom Office: https://widener.zoom.us/my/sevanbramer
>
>
>
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2037): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2037__;!!PvDODwlR4mBZyAb0!SsHphVJAk0TVUhOyHtyiLq0osbdU3h=
OQ7IrJLLl_IFsFjyzkMVKM4gzEpBomlqKCpBV1gXteIiqFw_jaVWG0Uv9GSIYQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/111187650=
/7559972__;!!PvDODwlR4mBZyAb0!SsHphVJAk0TVUhOyHtyiLq0osbdU3hOQ7IrJLLl_IFsFj=
yzkMVKM4gzEpBomlqKCpBV1gXteIiqFw_jaVWG0UsL8A9Q3$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Feb 14 2025 - 11:51:02 MST