Re: [AMMRL] problems with pulsecal --> One way valves for helium exhaust

From: Bob Berno <bberno_at_mcmaster.ca>
Date: Wed, 21 Feb 2024 14:31:59 -0500

Hello David,

Thank you for sharing your first hand experience with the Z58140 one-way valve.
This particular one-way valve I showed in the photograph was in the big
magnet installation crate that was delivered ahead of the magnet
rebuilding.  When I saw the "for installation only" stamped on the side
of valve, I had assumed it was just their way to ensure it didn't get
left behind by mistake when the work was done.

Now I see that there was a much more important reason that message was
stamped on the side of the valve.

Very good to know!

Cheers,
Bob.


On 2024-02-21 1:21 p.m., Vander Velde, David via groups.io wrote:
>
>
> Regarding the valve Z58140, I have some history with this ... above
> the part number, it says Nur fur installation, aka "only for
> installation." I should have taken that injunction at face value. I
> bought a couple of these figuring I could use them to connect some
> Oxford magnets to our helium recovery system, and leave them in place
> both during fills and between fills, like I did with Bruker magnets
> and Bruker check valves. However, this did not work in practice. At
> normal between-fill outflow rates, the gas flow on my flow meter would
> seesaw wildly every couple of seconds. It was not a good look. I think
> this valve is designed for popping open under much higher than normal
> flow, and staying open only when that flow and back pressure is
> sustained. Probably optimized for the conditions encountered during
> installing the magnet. So they cost me ca $800 each, but not fit for
> my intended purpose. Caveat emptor.
> ------------------------------------------------------------------------
> *From:* main_at_ammrl.groups.io <main_at_ammrl.groups.io> on behalf of Bob Berno <bberno_at_mcmaster.ca>
> *Sent:* Wednesday, February 21, 2024 10:04 AM
> *To:* main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> *Subject:* Re: [AMMRL] problems with pulsecal --> One way valves for helium exhaust
> Hello Karel,
>
> I can't comment on the main topic of your email, but I wanted to answer
> your questions about the one-way valve on your helium exhaust port.
> I don't know what the fittings look like at the top of your helium
> manifold, but from the photo you provided I suspect you have some sort
> of adapter that allows you to connect a piece of Tygon (or similar)
> tubing to the check valve you showed in your photo.
> Basically, you want to replace the adapter+small check valve with
> something like the silver one-way valve shown in the attached photo.
> Fortunately, this photo also shows the Bruker part number, for your
> ordering convenience.  About four years ago it cost about $365.00 CDN.
>
> This silver one-way valve is what we have on all but one of our
> magnets.  However, you might not be able to attach this directly to your
> existing manifold because it may be looking for something which has a
> KF-25 type fitting.  Then, you would also need a KF 25 adapter that will
> allow you to screw one-way valve into place.  Off hand I don't seem to
> have the part number for that.
>
> One other option would be to get a different one-way valve which already
> had the KF-25 fittings on both ends.  A few years ago when Bruker was on
> site to re-energize an old magnet, the magnet engineer installed a
> one-way valve like the one in the photo shown in this link:
> https://mcloud.mcmaster.ca/s/SM6963SwWSof4WK
> It's a little hard to see, but it looks like it has Part Number: Z58140
> I have no idea what this costs.
>
> I hope this helps.
> Bob.
>
--
Dr. Bob Berno
Manager, NMR Facility
McMaster University, ABB-156
1280 Main St. West
Hamilton, ON L8S 4M1
T: 905-525-9140 x24158 or x24686
---------------------------------------------------------------------------
This email contains CONFIDENTIAL information intended for the specific recipient(s).
If you have an accommodation need for a planned meeting, please e-mail me directly and
I will do my best to make appropriate arrangements.
Should you require any materials sent via this e-mail address in an alternate/accessible
format, please let me know.

McMaster University sits on the traditional Territories of the Mississauga and
Haudenosaunee Nations, and within the lands protected by the "Dish With One Spoon"
wampum agreement.

  


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#999): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/999__;!!PvDODwlR4mBZyAb0!UF69vcfVOP57JWY_u7OmnGY9L8wwZwcy=
T1-BwnEET1yuGZR2QyucOMSkg-C3Yq541wfhb4RQ-W1zEryc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104493541=
/7559972__;!!PvDODwlR4mBZyAb0!UF69vcfVOP57JWY_u7OmnGY9L8wwZwcyT1-BwnEET1yuG=
ZR2QyucOMSkg-C3Yq541wfhb4RQ-UgKMDO4$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Feb 21 2024 - 11:32:07 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 29 2024 - 14:59:21 MST