From: Karel Klika <karkli_at_utu.fi>
Date: Wed, 21 Feb 2024 15:25:07 +0000

Dear all,

We have a problem with pulsecal, sometimes it works, sometimes more-or-less
(i.e. poorly), sometimes really badly (10x or even 100x longer pulse width
than it should be), and sometimes fails completely (“no peak found in the
area”). The results can vary widely on immediate consecutive runs of pulsecal,
e.g. a good or near reasonable result to complete failure or a nonsense result.
Pulsecal has been used for many years with no problems, now suddenly not.
Attached are pics of the nutation when it fails completely. I have run ha for
the amplifier and it seems fine, the pulse widths for proton (both hard and soft)
and an X nucleus are fine, both decoupling and pre-saturation for proton are
also working as per normal, and sensitivities for both proton and an X nucleus
also seem fine. The probe is a 5-mm BBO (Z108618_0274) and the console is an
Avance III (H03128OB). I have rebooted TopSpin (version 3.6.2 running under
Windows 7) and the console multiple times, recompiled pulsecal, and used the
pulse sequence pulsecal in place of pulsecal.2. Usually, pulsecal seems to work
with just about any sample, but I have been testing on the standard sucrose in
90% H2O. What could be wrong and how to fix the problem ? Let me know please
what I should try to diagnose the fault/fix the problem.

I don’t know if this is related, but usually the nutation arrays (popt) that
I run are fine, but this one result (pic attached) is very strange, and it
was run immediately after a normal looking result on the same sample for the
same pulse width determination (just run with smaller steps for finer evaluation).
Subsequent runs on other samples were also normal looking. The strange result
is not the result of too high an rg value as it was not changed. It even
resembles what a double FT of the signals might look like ?


Also,

1. Anyone know what is the heat output on the various low, medium, and high
settings for a BCU-II ? It is listed as 2.4 kW, but would it be fair to assume
that this is for the high setting and that for the low setting it is only 0.5 kW,
i.e. the same as the BCU-I ?

2. Is it possible to fit a modern helium check valve to our 2010 vintage magnet ?
(pics attached) When we do helium fills we take the current check valve off; in
my previous post we left the more modern check valve in place. So can we do helium
fills with the modern check valve in place on this magnet if we can replace it ?
Anyone happen to know the part number for the more modern check valve ?

Thank you very much in advance.

Kind regards,
Karel Klika


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#994): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/994__;!!PvDODwlR4mBZyAb0!UmsZ4lMG2dH3a3CJMduJQo62PtArYxEI=
2k_uHjEAZGO-dorOyUF-beNuXUlcK3XCBx7mTu0n8tCxaA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104490589=
/7559972__;!!PvDODwlR4mBZyAb0!UmsZ4lMG2dH3a3CJMduJQo62PtArYxEI2k_uHjEAZGO-d=
orOyUF-beNuXUlcK3XCBx7mTu38ojHNfw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-





IMG_0003.jpeg
(image/jpeg attachment: IMG_0003.jpeg)

IMG_0004.jpeg
(image/jpeg attachment: IMG_0004.jpeg)

IMG_0005.jpeg
(image/jpeg attachment: IMG_0005.jpeg)

example1 PW x 10 true value.png
(image/png attachment: example1_PW_x_10_true_value.png)

example2 no peak found.png
(image/png attachment: example2_no_peak_found.png)

popt.png
(image/png attachment: popt.png)

Received on Wed Feb 21 2024 - 07:46:59 MST

This archive was generated by hypermail 2.4.0 : Thu Feb 29 2024 - 14:59:21 MST