Hi John,
This really sounds to me like something got hosed up with the routing diagrams.
Would you mind sending pictures of the diagram you get with edprobe and also
the edasp diagram with “show receiver wiring” checked?
Cheers,
Mike
On Feb 8, 2024, at 11:20 AM, John H Grimes Jr. <jgrimesjr_at_uga.edu> wrote:
Greetings Spinlanders,
We have an Avance Neo 600 NMR running TopSpin 4.1.4. We switch between probes
on this system and recently when we switched from a solids probe to a liquids
probe, we lost ability to view a wobble curve (we just see noise). The symptom
are as follows. With a BBO probe installed and routing through a HPPR2 XBB19F
2HP slice, we can only see noise when we wobble 13C (or any other X nucleus),
however we can acquire a 13C spectrum. If we switch the cable from the probe
to the preamp such that the signal goes through a HPPR/2 13C module and adjust
the routing in the software (edprobe & edasp), we can view a wobble trace but
can’t obtain a 13C spectrum. As part of this switching process I’ve
rebooted the spectrometer, run cf, run expinstall, power cycled the slices by
logging into the blade epu and using the pdu. I’ve noticed that when I
use expinstall some parameter sets aren’t installed (e.g. C13CPD).
There are no errors reported in the output from cf, nor that I see from expinstall.
The proton channel works fine. Our hypothesis is there is a file that is missing
or is corrupted and this was possibly due to expinstall. Without resorting to
upgrading to a new installation of TopSpin or backtracking to a several month old
nmrsaved version, what files could be missing or possible corrupt to cause this
behavior?
Thanks,
John
John Grimes
Complex Carbohydrate Research Center
University of Georgia
315 Riverbend Road, Athens, GA 30602
(706) 542-4483
jgrimesjr_at_uga.edu
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#978): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/978__;!!PvDODwlR4mBZyAb0!X_arM6hJiwfGYPyHsNoCcZmUDs0Ig8Kf=
SxUo-pFbR3vHNO1EN9Rieq8VVaU0Sc6nj5GsWQZnRQVB0a_7Hw$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/104241640=
/7559972__;!!PvDODwlR4mBZyAb0!X_arM6hJiwfGYPyHsNoCcZmUDs0Ig8KfSxUo-pFbR3vHN=
O1EN9Rieq8VVaU0Sc6nj5GsWQZnRQVGDX-Cww$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Feb 08 2024 - 09:52:00 MST