Hello Everyone,

Thank you all for your support!

We could fix the problem with your tips. The capacitors of the green box
were fried and after replacing them (0.5 € per piece) the
spectrometer is running again.

Kind regards,

Jörn

> Von: main_at_ammrl.groups.io Im Auftrag von joern.droste via groups.io
> Gesendet: Montag, 6. Oktober 2025 15:22
> An: main_at_ammrl.groups.io
> Betreff: [extern] [AMMRL] Bruker Avance I Power outage #Bruker #Hardware

Hello Everyone,

We recently experienced a brief power outage lasting only some seconds or up
to a minute.

Since then, our Bruker Avance I spectrometer has not been operational.

Specifically, the cabinet does not receive power when I attempt to switch it
on using the blue button at the cabinet.

After reviewing past discussions on this mailing list, I've come to suspect
that a fuse might be the cause of the issue.

I believe the problem could be related to either the green box or the ESB
63-40 fuse (please see the attached picture for reference).

Can anyone kindly offer guidance on how I might verify if a fuse is indeed
the problem, and if so, how to repair it?

Any advice or insights would be greatly appreciated.

Thank you in advance for your support.

Kind regards

Jörn



Dr. Jörn Droste
NMR-Abteilung des Instituts für Organische Chemie
Leibniz Universität Hannover
Schneiderberg 1B
30167 Hannover
Tel.: +49-(0)511-762-4644
Tel.: +49-(0)511-762-4647


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3051): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3051__;!!PvDODwlR4mBZyAb0!UAvx20hJhgL0NqCBaGDLxTVYFmqHb3=
29vCsQv5owGfxnrsRHY4h036cWiuMMr2glD7WnjBB4cNkIBMgfrNt6QDYxj6e3$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/116792697=
/7559972__;!!PvDODwlR4mBZyAb0!UAvx20hJhgL0NqCBaGDLxTVYFmqHb329vCsQv5owGfxnr=
sRHY4h036cWiuMMr2glD7WnjBB4cNkIBMgfrNt6QImXO6bs$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!UAvx20hJhgL0NqCBaGDLxTVYFmqHb329vC=
sQv5owGfxnrsRHY4h036cWiuMMr2glD7WnjBB4cNkIBMgfrNt6QGI6GwNW$
Mute #bruker:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mut=
ehashtag/bruker__;!!PvDODwlR4mBZyAb0!UAvx20hJhgL0NqCBaGDLxTVYFmqHb329vCsQv5=
owGfxnrsRHY4h036cWiuMMr2glD7WnjBB4cNkIBMgfrNt6QEVOzbYe$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Dec 15 2025 - 06:41:04 MST

This archive was generated by hypermail 2.4.0 : Thu Jan 01 2026 - 15:24:06 MST