Hi Raj,
There are a couple of things you might try. I would use the ha command
in topspin and see if you can access the BSMS service webpage. Or even
open a terminal/command prompt and just try pinging 149.236.99.20 or 149.236.99.21.
If you can communicate with the ELCB you might be able to either reset it or to
find out what the problem is.
And this could also be related to a problem with the keyboard. I would
suggest powering off the BSMS rack, disconnecting the keyboard cable from the
ELCB, and powering back up to see if it will boot up. As much as people
love these keyboards, they do get old after a while and can fail.
Cheers,Mike
On Monday, December 9, 2024 at 11:33:21 AM MST, Rajendran Kaliaperumal via groups.io wrote:
Hello MembersWe have Oxford 400 MHz with Bruker Avance II console started giving
Error 70 - BSMS timeout connecting error. A month ago, it was having this issue
and left it for connecting. For our surprise, it automatically connected and
solved the error. Now it started again and not connecting at all. I can see the
LED red indications in all the three Boards (SLCB/2, GAB, SCB20) of BSMS/2 as
seen in the photos attached. I tried removing these boards and re-seated them.
Nothing helped. I looked the LEDs in the Back of BSMS, all solid green. I checked
the voltage selector (correct voltage selected). The same error happened in 2022
and ended up changing ELCB board as Bruker recommended. There was some power
outages happened before the error pop-ups. However, we have UPS to control the
voltage and restart. So, there was no expectation of impact of power surges.The
error RED in SGU400 is normal and it will be cleared once the power cycle was
finished.The RED error in SLCB/2, GAB, SCB20 are showing because of error 70: BSMS
not connecting .... I couldn't do the power cycle to get into 'ii' with Topspin,
because of the BSMS error. Looking for your expert advice on this issue -
what further I can check up to diagnose the problem.
With warm regards
Raj Kaliaperumal, PhD
Instrument Tech./SpecialistDepartment of Chemistry
Cape Breton University
Sydney, NS
CAraj_kalia_at_cbu.ca
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1922): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1922__;!!PvDODwlR4mBZyAb0!X3_R0BRPG4HYQjU-vy2_2ATrB1geyg=
tpcjodLwE_x_oSt7cJvFIefyVp-CWvBsh7bT5WLE1cRUtvoIEkApXa3Cyn1gRt$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/110011617=
/7559972__;!!PvDODwlR4mBZyAb0!X3_R0BRPG4HYQjU-vy2_2ATrB1geygtpcjodLwE_x_oSt=
7cJvFIefyVp-CWvBsh7bT5WLE1cRUtvoIEkApXa3JT9J6WJ$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Mon Dec 09 2024 - 12:08:15 MST