From: Holmes, Daniel <holmesd5_at_msu.edu>
Date: Tue, 5 Dec 2023 14:06:36 +0000

An alternative and addition to Clemens’ approach if you add the
following script to the Title options in MNova, you’ll get the
experiment type on the spectra.

{parm,"Experiment"}

> From: main_at_ammrl.groups.io on behalf of Clemens Anklin via groups.io
> Date: Tuesday, December 5, 2023 at 8:39 AM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Bruker: Experiment numbers - Experiment Names?

Dear Ken

to import your data into MNova you could indeed use the following structure.

In a folder with the username you make a folder for the "sample" , then you
use the parameter NAME for the experiment name and always use EXPNO = 1.

If you drag the folder "samplename" in MNova it will open all experiments in
one document.

Maybe this helps

Clemens

On 12/4/2023 2:36 PM, Kenneth Sharp-Knott wrote:

We have a few groups who create a new filename for every experiment resulting
in numerous folders for the same sample. I find this sloppy and frustrating.

What is the typical procedure at your institution for getting around Brukers
numbering system which doesn't provide the experiment name? Is there a way
to include more than just the number of the experiment in the sample folder?

--
Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)

-- =========================
==========================
=====================
Clemens Anklin Ph.D. Vice President Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667) Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
======================
==========================
========================



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#835): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/835__;!!PvDODwlR4mBZyAb0!TGRvgRw5cBIefIDz5Csz_bjKaY12_xIr=
JCbKaDgalVaTH-J-cAGxY6xSZqhfLjebA2vEyYDMibWlko1p$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102977754=
/7559972__;!!PvDODwlR4mBZyAb0!TGRvgRw5cBIefIDz5Csz_bjKaY12_xIrJCbKaDgalVaTH=
-J-cAGxY6xSZqhfLjebA2vEyYDMiYNzs6_X$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Dec 05 2023 - 06:11:29 MST

This archive was generated by hypermail 2.4.0 : Mon Jan 01 2024 - 19:06:37 MST