Re: [AMMRL] Bruker: Experiment numbers - Experiment Names?

From: Clemens Anklin via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Tue, 5 Dec 2023 02:10:12 +0000

Dear Ken

to import your data into MNova you could indeed use the following structure.

[cid:part1.1yoGz2lD.Vxx78xRg_at_bruker.com]

In a folder with the username you make a folder for the "sample" , then you
use the parameter NAME for the experiment name and always use EXPNO = 1.

If you drag the folder "samplename" in MNova it will open all experiments in
one document.

Maybe this helps

Clemens

On 12/4/2023 2:36 PM, Kenneth Sharp-Knott wrote:

We have a few groups who create a new filename for every experiment resulting
in numerous folders for the same sample. I find this sloppy and frustrating.

What is the typical procedure at your institution for getting around Brukers
numbering system which doesn't provide the experiment name? Is there a way
to include more than just the number of the experiment in the sample folder?

--
Ken Sharp-Knott
Manager of Analytical Services and the NMR Facility
Department of Chemistry
Virginia Tech

(540)267-6502 (Cell)
(540)231-0885 (Office)



--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#833): https://urldefense.com/v3/__https://ammrl.groups.=
io/g/main/message/833__;!!PvDODwlR4mBZyAb0!VdRpvc2LJwIbHIS10xrDw3h9SSolnRfB=
M0QgzPk_KgBJx3GIUC6206NghLHOqcsTR2dTyL7Kc6cSgO-87rn94KqBRL06$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/102977754=
/7559972__;!!PvDODwlR4mBZyAb0!VdRpvc2LJwIbHIS10xrDw3h9SSolnRfBM0QgzPk_KgBJx=
3GIUC6206NghLHOqcsTR2dTyL7Kc6cSgO-87rn94D0MiLbp$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-






qcJAkiDXARO0ZqMw.png
(image/png attachment: qcJAkiDXARO0ZqMw.png)

Received on Tue Dec 05 2023 - 05:39:46 MST

This archive was generated by hypermail 2.4.0 : Mon Jan 01 2024 - 19:06:37 MST